Trace Oxygen Analyzer | Installation | |
|
|
|
2.2.4 The Effect of Pressure
In order to state the amount of oxygen present in the sample as a percentage of the gas mixture, it is necessary that the sample diffuse into the cell under constant pressure.
If the total pressure increases, the rate that oxygen reaches the cathode through the diffusing membrane will also increase. The electron transfer, and therefore the external current, will increase, even though the oxygen concentration of the sample has not changed. It is therefore important that the sample pressure at the fuel cell (usually vent pressure) remain constant between calibrations.
2.2.5 Calibration Characteristics
Given that the total pressure of the sample gas at the surface of the
In addition, since there is zero output in the absence oxygen, the characteristic curve is close to an absolute zero. The cell itself does not need to be zeroed on ranges as sensitive as
As the cell reaches the end of its useful life, the slope seen in Figure