Installation

 

 

 

2.2.4 The Effect of Pressure

In order to state the amount of oxygen present in the sample as a percentage of the gas mixture, it is necessary that the sample diffuse into the cell under constant pressure.

If the total pressure increases, the rate that oxygen reaches the cathode through the diffusing membrane will also increase. The electron transfer, and therefore the external current, will increase, even though the oxygen concentration of the sample has not changed. It is therefore important that the sample pressure at the fuel cell (usually vent pressure) remain constant between calibrations.

2.2.5 Calibration Characteristics

Given that the total pressure of the sample gas at the surface of the Micro-fuel Cell input is constant, a convenient characteristic of the cell is that the current produced in an external circuit is directly proportional to the rate at which oxygen molecules reach the cathode. This rate is directly proportional to the concentration of oxygen in the gaseous mixture. In other words it has a linear characteristic curve, as shown in Figure 2-2. Measuring circuits do not have to compensate for nonlinearity.

In addition, since there is zero output in the absence oxygen, the characteristic curve is close to an absolute zero. The cell itself does not need to be zeroed on ranges as sensitive as 0-10 ppm. The electronics are zeroed to compensate for zero offsets in the electronics. (The electronics is zeroed automatically when the instrument power is turned on during the Self-Test.)

As the cell reaches the end of its useful life, the slope seen in Figure 2-2 decreases. As this occurs, the span adjustments will become larger.

Teledyne Analytical Instruments

9

Image 17
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 2000 Teledyne Analytical Instruments Model OT-3Specific Model Information Important Notice Trace Oxygen Analyzer Table of Contents Appendix MaintenanceOperation Front Panel List of FiguresIntroduction Main Features of the AnalyzerOverview Trace Oxygen Analyzer IntroductionIntroduction OT-3 Front Panel DescriptionFunction Keys Measurement Parameters Readout Rear Panel DescriptionRear Panel AC Version Operational Theory Installation OT-3Introduction Micro-fuel Cell SensorAnatomy of a Micro-fuel Cell Trace Oxygen Analyzer InstallationElectrochemical Reactions Effect of Pressure Calibration CharacteristicsGeneral ElectronicsBlock Diagram of the Signal Processing Electronics Signal ProcessingInstallation OT-3 Unpacking the Analyzer InstallationElectrical Connections Installing the Micro-fuel CellLocation and Mounting Analog Outputs Primary Input PowerContact ID for Failsafe Relay Operation Alarm RelaysRelay Ratings Solid State Relay Output Gas ConnectionsGas Connections Sample System OT-3 Sample System Installation ChecklistInstallation OT-3 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation OperationOperation OT-3 Using the Function and Data Entry ButtonsSetting the Analysis Ranges Set Alarm Setting the Alarm SetpointsHI Range LO RangePower Failure Alarm Alarm DelayKeypad Lockout Calibration Selecting a Fixed Range or AutorangingSupplementary Information Displaying Percent & PPM on the LED DisplayOperation OT-3 Replacing the Fuse MaintenanceTrace Oxygen Analyzer Maintenance When to Replace a Sensor Sensor Installation or ReplacementMaintenance OT-3 Removing the Micro-fuel Cell Ordering and Handling of Spare SensorsInstalling a Micro-fuel Cell Cell Warranty Conditions AppendixOT-3 SpecificationsAppendix Analytical Instruments Spare Parts ListTrace Oxygen AnalyzerAppendix Appendix OT-3 Reference DrawingMiscellaneous Trace Oxygen Analyzer Index IndexNo Pass Index OT-3Website address Index OT-3 Teledyne Analytical Instruments