Maintenance |
|
|
|
Figure 5-1: AC Fuse Replacement
DC Units
On DC units, the fuse is located in a fuse holder on the rear panel. See Figure
5.2Sensor Installation or Replacement
5.2.1 When to Replace a Sensor
On trace analyzers the
To do this, you should connect a pressure regulated zero gas source to the instrument at the span gas port. This can be done by either shutting off the span gas and replacing it with the zero gas supply or by teeing into the span gas line using a selector valve which can then feed either span or zero gas to the instrument.
It may take several hours for the sample system and cell to return to a stable low level of O2 if it has been exposed to higher levels of O2. If the instrument will not come down to an acceptably low reading on zero gas, then a depleted cell or a sample system leak should be suspected.
Typically offsets of 1.0 ppm or less are acceptable. A cell failure may also be indicated by an inability to SPAN, or slow response to changes in O2 concentration at levels below 100 ppm.
Teledyne Analytical Instruments | 32 |