Teledyne OT-3 operating instructions Primary Input Power, Analog Outputs

Page 23

Trace Oxygen Analyzer

Installation

 

 

 

Figure 3-1: Electrical Connectors for AC Control Unit

3.3.1 Primary Input Power

A detachable power cord is installed directly into the Control Unit and connects to a three pin terminal block on the rear panel. The terminal block is provided for customer AC wiring.

The universal power supply allows direct connection to any 100-240 VAC, 50/60Hz power source. The fuse block, to the right of the power cord receptacle, accepts two 5x20mm 0.5 A, 250V, IEC time-lag (T) fuse. (See Fuse Replacement in chapter 5, Maintenance.)

For DC powered units, the customers DC power wiring is attached to the two pin terminal block located on the rear of the control unit. See Figure 1-3.

3.3.2 Analog Outputs

There are three DC (four on AC units) output signal connectors with screw terminals on the panel. There are two wires per output with the polarity noted. See Figure 3-1. The outputs are:

0–10 V % Range: Voltage rises with increasing oxygen concentration, from 0 V at 0 percent oxygen to 10 V at full scale percent oxygen. (Full scale = 100% of programmed range.)

0–10 V Range ID: 3.33 V = Low Range, 6.66 V = High Range, 10 V = Air Cal Range.

Teledyne Analytical Instruments

15

Image 23
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 2000 Teledyne Analytical Instruments Model OT-3Specific Model Information Important Notice Trace Oxygen Analyzer Table of Contents Appendix MaintenanceOperation Front Panel List of FiguresTrace Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction OverviewIntroduction OT-3 Front Panel DescriptionFunction Keys Measurement Parameters Readout Rear Panel DescriptionRear Panel AC Version Micro-fuel Cell Sensor Installation OT-3Operational Theory IntroductionAnatomy of a Micro-fuel Cell Trace Oxygen Analyzer InstallationElectrochemical Reactions Effect of Pressure Calibration CharacteristicsGeneral ElectronicsBlock Diagram of the Signal Processing Electronics Signal Processing Installation OT-3 Unpacking the Analyzer InstallationElectrical Connections Installing the Micro-fuel CellLocation and Mounting Analog Outputs Primary Input PowerContact ID for Failsafe Relay Operation Alarm RelaysRelay Ratings Solid State Relay Output Gas ConnectionsGas Connections Sample System OT-3 Sample System Installation ChecklistInstallation OT-3 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation OperationOperation OT-3 Using the Function and Data Entry ButtonsSetting the Analysis Ranges LO Range Setting the Alarm SetpointsSet Alarm HI RangePower Failure Alarm Alarm DelayKeypad Lockout Calibration Selecting a Fixed Range or AutorangingSupplementary Information Displaying Percent & PPM on the LED DisplayOperation OT-3 Replacing the Fuse MaintenanceTrace Oxygen Analyzer Maintenance When to Replace a Sensor Sensor Installation or ReplacementMaintenance OT-3 Removing the Micro-fuel Cell Ordering and Handling of Spare SensorsInstalling a Micro-fuel Cell Cell Warranty Conditions AppendixOT-3 SpecificationsAppendix Analytical Instruments Spare Parts ListTrace Oxygen AnalyzerAppendix Appendix OT-3 Reference DrawingMiscellaneous Trace Oxygen Analyzer Index IndexNo Pass Index OT-3Website address Index OT-3 Teledyne Analytical Instruments