Omega 27, 25, 26, 24 manual Memory Backup, Operation with Rechargeable Battery

Page 12

2.Remove the old battery by lifting the battery extractor loop.

3.Place the new battery in the battery compartment. Be sure to observe proper polarity.

4.Re-install the battery cover before resuming use of the instrument.

NOTES:

Less than 10% of battery life remains when the BAT annuncia- tor turns on.

When the battery is excessively discharged, the display is blanked.

If the instrument is going to be stored for a long period of time or in a high temperature environment, remove the battery to prevent leakage damage.

After a new battery is installed, allow approximately 5 sec- onds for display turn-on the first time the unit is turned on. During this period, microcomputer initialization is per- formed.

MEMORY BACKUP

During battery replacement, the contents of user-programmed memory (data, operating modes, etc.) can be saved. Prior to removing the old battery, turn off the instrument, and connect an HH22-AC battery charger. Then exchange batteries, and disconnect the battery charger. Do not leave the bat- tery charger connected to instruments with non-rechargeable batteries.

OPERATION WITH RECHARGEABLE

BATTERY

HH22-ACprovides a 9-volt Ni-Cd battery and recharger suitable for use with the unit. This battery provides 3-4 hours of continuous operation. This dura- tion can be extended to several days by operating simultaneously from both the battery and recharger.

Turn off the instrument to recharge the battery. Allow 7 hours to fully recharge the battery. Then unplug the recharger to avoid overcharging.

10

Image 12
Contents CL23A, 24, 25, 26 Omega.comTM Table of Contents Common Specifications General InformationTemperature Coefficient Page Unique Specifications Model CL23A/CL23ECModel CL24 Model CL25/CL25ECFeatures Manual AddendaModel CL27/CL27EC Safety Information Safety PrecautionsUnpacking Your Unit Page Battery INSTALLATION/REPLACEMENT Battery InstallationMemory Backup Operation with Rechargeable BatteryDiagnostics and Error Messages Problem Likely CauseOperating Instructions Initial Turn onKeypad Meter Mode Operation Thermocouples Set-Up OPRMeter Mode Operation RTD, Thermistor & Ohms Set-Up HoldData Logging and Recall To Save Readings To Recall ReadingsTo Erase Data Locations Meter Mode Calibrator Mode Operation Thermocouples Set-UpCalibrator Mode Operation RTD, Thermistor & Ohms Set-Up CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTERWire Resistance Simulation Storage and Recall of Calibrator Settings To Store OutputsTo Erase Set-Up Locations Calibrator Mode To Recall OutputsStep Function Ramp Function CL25 and CL27Service Information Calibration ProceduresThermocouple Calibration Test Equipment Required Ambient ConditionsPage Calibration Adjustments Meter-Mode Zero & Gain Adjustments Sensor Step Input Adjust ReadingCalibrator-Mode Zero & Gain Adjustments Sensor Output Adjust ReadingCalibration Setup, Copper Mode Resistance Calibration Test Equipment RequiredResistance Calibration Setup Calibration Adjustments Output Input Adjust Verify Reading Input Adjust Verify ReadingCalibration of Meter-Mode Resistance Functions RTD-Thermistor Sensor Unit StepCalibration Verification Model Software Revision Equipment RequiredPage Page Alloy Mode Verification Procedure for Thermocouple Functions Page Mechanical Parts Diagram Calibration Procedures Page Page Page Page Page USAWARRANTY/DISCLAIMER Temperature