Omega 26, 27, 25, 24 manual

Page 6

KEYPAD: 12 momentary switches with tactile feedback select;

On/Off

Change/Enter

Clear/–0

Sensor Select/1

°F/°C/2

Resolution/3

Operate/4

Store/5

Recall/6

Calibrator/Meter/Up Ramp/7

Calibrator Step/Down Ramp/8

Calibrator Ramp/Meter Hold/9

DISPLAY: 5 digit LCD, 0.4" height, and decimal point. Annunciators;

Loaded Memory Locations (10 for data storage, 4 for ramp parameters)

Low Battery (BAT)

Number Change Mode (NUM)

Calibrator Mode (CALIB)

Meter Mode (METER)

Operate Mode (OPR)

Store Mode (STO)

Recall Mode (RCL)

Meter Hold (HOLD)

Scale (°F/°C)

Sensor Type

POWER OFF CONFIGURATION RETENTION: Instrument retains last selected;

Sensor Type

°F/°C Scale

0.1°/1° Resolution

Calibrator/Meter Mode

Memory Contents (data storage and ramp parameters).

DIAGNOSTICS: Display codes indicate following conditions;

'BAT': Low Battery

'OPEN': Open Input Thermocouple, RTD or Thermistor.

'E-1' (momentary): Invalid Keypad Entry

'E-2' (momentary): Ramp Function Not Installed

'HI': Temperature Reading or Keypad Entry Exceeds Rating

'LO': Temperature Reading or Keypad Entry Too Low For Accuracy

LCD Test: During power-up, all segments and annunciators enabled momentarily

ENVIRONMENTAL LIMITS FOR OPERATING: 0°C to 50°C, less than 90% relative humidity (R.H.) up to 35°C; reduce R.H. limit by 3%/°C from 35°C to 50°C.

ENVIRONMENTAL LIMITS FOR STORAGE: –35°C to 60°C, less than 95% relative humidity (R.H.) up to 35°C; reduce R.H. limit by 3%/°C from 35°C to 60°C.

POWER: 9 volt alkaline battery (NEDA 1604A).

4

Image 6
Contents CL23A, 24, 25, 26 Omega.comTM Table of Contents Common Specifications General InformationTemperature Coefficient Page Model CL24 Unique SpecificationsModel CL23A/CL23EC Model CL25/CL25ECFeatures Manual AddendaModel CL27/CL27EC Safety Information Safety PrecautionsUnpacking Your Unit Page Battery INSTALLATION/REPLACEMENT Battery InstallationMemory Backup Operation with Rechargeable BatteryOperating Instructions Diagnostics and Error MessagesProblem Likely Cause Initial Turn onKeypad Meter Mode Operation Thermocouples Set-Up OPRMeter Mode Operation RTD, Thermistor & Ohms Set-Up HoldData Logging and Recall To Save Readings To Recall ReadingsTo Erase Data Locations Meter Mode Calibrator Mode Operation Thermocouples Set-UpCalibrator Mode Operation RTD, Thermistor & Ohms Set-Up CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTERWire Resistance Simulation Storage and Recall of Calibrator Settings To Store OutputsTo Erase Set-Up Locations Calibrator Mode To Recall OutputsStep Function Ramp Function CL25 and CL27Thermocouple Calibration Test Equipment Required Service InformationCalibration Procedures Ambient ConditionsPage Calibration Adjustments Calibrator-Mode Zero & Gain Adjustments Sensor Meter-Mode Zero & Gain Adjustments Sensor StepInput Adjust Reading Output Adjust ReadingCalibration Setup, Copper Mode Resistance Calibration Test Equipment RequiredResistance Calibration Setup Calibration Adjustments Calibration of Meter-Mode Resistance Functions Output Input Adjust Verify ReadingInput Adjust Verify Reading RTD-Thermistor Sensor Unit StepCalibration Verification Model Software Revision Equipment RequiredPage Page Alloy Mode Verification Procedure for Thermocouple Functions Page Mechanical Parts Diagram Calibration Procedures Page Page Page Page Page USAWARRANTY/DISCLAIMER Temperature