NOTE: For calibrators with the older, single-jumper configura- tion, refer to calibration procedure beginning on page 39.

14. Install calibration cover in place of bottom-cover.

15. Re-install battery.

16. Hook up test-equipment, calibration cables, and U.U.T. per Figure 10.

17. Turn on DMM and TC simulator for warmup. Allow at least 30 minutes.

18. Set the DMM to 200mV DC range.

19. Set the TC simulator to its COPPER output mode, with a reference tem- perature of 32.0°F.

10.Turn on U.U.T. and verify that "RTD" and "Ω " annunciators are flashing on the display. This verifies removal of J1 and J2 respectively.

NOTE: Flashing “RTD” indicates that zero-offset corrections can be stored in EEPROM. It also indicates that cold-junction com- pensation is disabled.

Flashing 'Ω ' indicates that error corrections for the cold-junction sensors can be stored or recalled from EEPROM.

11.Clear the EEPROM locations used to store E1 (meter cold-junction sen- sor error) and E2 (calibrator cold-junction sensor error) as follows:

a.Set U.U.T to CALIB, OPR, K, °F mode.

b.Key in: CHANGE/0.0/ENTER/STO/1/OPR (E1 is set to 0.0°F)

c.Key in: CHANGE/0.0/ENTER/STO/2/OPR (E2 is set to 0.0°F)

NOTE: Both 1 & 2 memory annunciators should be turned on, but the “RTD” annunciator will have turned off.

d.Perform RECALL 1 and RECALL 2 to verify storage of 0.0°F in both locations.

NOTE: At the completion of calibration, E1 and E2 will be saved in EEPROM. Memory locations 1 and 2 will be free for normal use.

e.Set U.U.T. to METER, OPR, K, °F.

f.Turn U.U.T. OFF, then ON to re-enable 'RTD' andΩ ' annunciators.

24

Image 26
Contents CL23A, 24, 25, 26 Omega.comTM Table of Contents Common Specifications General InformationTemperature Coefficient Page Model CL24 Unique SpecificationsModel CL23A/CL23EC Model CL25/CL25ECModel CL27/CL27EC FeaturesManual Addenda Unpacking Your Unit Safety InformationSafety Precautions Page Battery INSTALLATION/REPLACEMENT Battery InstallationMemory Backup Operation with Rechargeable BatteryOperating Instructions Diagnostics and Error MessagesProblem Likely Cause Initial Turn onKeypad Meter Mode Operation Thermocouples Set-Up OPRMeter Mode Operation RTD, Thermistor & Ohms Set-Up HoldData Logging and Recall To Save Readings To Recall ReadingsTo Erase Data Locations Meter Mode Calibrator Mode Operation Thermocouples Set-UpCalibrator Mode Operation RTD, Thermistor & Ohms Set-Up CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTERWire Resistance Simulation Storage and Recall of Calibrator Settings To Store OutputsTo Erase Set-Up Locations Calibrator Mode To Recall OutputsStep Function Ramp Function CL25 and CL27Thermocouple Calibration Test Equipment Required Service InformationCalibration Procedures Ambient ConditionsPage Calibration Adjustments Calibrator-Mode Zero & Gain Adjustments Sensor Meter-Mode Zero & Gain Adjustments Sensor StepInput Adjust Reading Output Adjust ReadingCalibration Setup, Copper Mode Resistance Calibration Test Equipment RequiredResistance Calibration Setup Calibration Adjustments Calibration of Meter-Mode Resistance Functions Output Input Adjust Verify ReadingInput Adjust Verify Reading RTD-Thermistor Sensor Unit StepCalibration Verification Model Software Revision Equipment RequiredPage Page Alloy Mode Verification Procedure for Thermocouple Functions Page Mechanical Parts Diagram Calibration Procedures Page Page Page Page Page USAWARRANTY/DISCLAIMER Temperature