NOTE: For calibrators with the older,
14. Install calibration cover in place of
15.
16. Hook up
17. Turn on DMM and TC simulator for warmup. Allow at least 30 minutes.
18. Set the DMM to 200mV DC range.
19. Set the TC simulator to its COPPER output mode, with a reference tem- perature of 32.0°F.
10.Turn on U.U.T. and verify that "RTD" and "Ω " annunciators are flashing on the display. This verifies removal of J1 and J2 respectively.
NOTE: Flashing “RTD” indicates that
Flashing 'Ω ' indicates that error corrections for the
11.Clear the EEPROM locations used to store E1 (meter
a.Set U.U.T to CALIB, OPR, K, °F mode.
b.Key in: CHANGE/0.0/ENTER/STO/1/OPR (E1 is set to 0.0°F)
c.Key in: CHANGE/0.0/ENTER/STO/2/OPR (E2 is set to 0.0°F)
NOTE: Both 1 & 2 memory annunciators should be turned on, but the “RTD” annunciator will have turned off.
d.Perform RECALL 1 and RECALL 2 to verify storage of 0.0°F in both locations.
NOTE: At the completion of calibration, E1 and E2 will be saved in EEPROM. Memory locations 1 and 2 will be free for normal use.
e.Set U.U.T. to METER, OPR, K, °F.
f.Turn U.U.T. OFF, then ON to
24