23.Select Type-E thermocouple.
24.Turn off U.U.T.
25.Replace T calibration-cable with E calibration-cable.
26.Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.
27.Apply 32°F (Type E). Adjust P1 for a reading of 32.0°F (exclusive of noise).
28.Hit OPR key to store E-meter zero-adjustment.
29.Select Type-K thermocouple.
30.Select CALIB mode.
31.Turn off U.U.T.
32.Connect U.U.T. to test-equipment per Figure 3 using the K calibration- cable.
33.Turn-on U.U.T. and allow 2-3 minutes, for thermal stabilization
34.Set both the U.U.T. and TC-simulator to output 32.0°F (Type K). Adjust P2 for a null-meter reading of 0±3µV.
35.Set both the U.U.T. and T-simulator to output 2400°F (Type K). Adjust P41 for a null-meter reading of 0±3µV.
NOTE: Steps 34 and 35 are interactive. Repeat each as neces- sary to get the desired null readings.
36.Hit OPR key to store K-calib zero adjustment.
37.Select J thermocouple.
38.Turn off U.U.T.
39.Replace K calibration-cable with J calibration-cable.
40.Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.
41.Set both the U.U.T. and TC-simulator to output 32.0°F (Type J). Adjust P1 for a null-meter reading of 0±3µV.
42.Hit OPR Key to store J-calib zero adjustment.
43.Select Type-T thermocouple.
44.Turn off U.U.T.
45.Replace J calibration-cable with T calibration-cable.
46.Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.