23.Select Type-E thermocouple.

24.Turn off U.U.T.

25.Replace T calibration-cable with E calibration-cable.

26.Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.

27.Apply 32°F (Type E). Adjust P1 for a reading of 32.0°F (exclusive of noise).

28.Hit OPR key to store E-meter zero-adjustment.

29.Select Type-K thermocouple.

30.Select CALIB mode.

31.Turn off U.U.T.

32.Connect U.U.T. to test-equipment per Figure 3 using the K calibration- cable.

33.Turn-on U.U.T. and allow 2-3 minutes, for thermal stabilization

34.Set both the U.U.T. and TC-simulator to output 32.0°F (Type K). Adjust P2 for a null-meter reading of 0±3µV.

35.Set both the U.U.T. and T-simulator to output 2400°F (Type K). Adjust P41 for a null-meter reading of 0±3µV.

NOTE: Steps 34 and 35 are interactive. Repeat each as neces- sary to get the desired null readings.

36.Hit OPR key to store K-calib zero adjustment.

37.Select J thermocouple.

38.Turn off U.U.T.

39.Replace K calibration-cable with J calibration-cable.

40.Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.

41.Set both the U.U.T. and TC-simulator to output 32.0°F (Type J). Adjust P1 for a null-meter reading of 0±3µV.

42.Hit OPR Key to store J-calib zero adjustment.

43.Select Type-T thermocouple.

44.Turn off U.U.T.

45.Replace J calibration-cable with T calibration-cable.

46.Turn on U.U.T. Allow 2-3 minutes for thermal stabilization.

41

Image 43
Contents CL23A, 24, 25, 26 Omega.comTM Table of Contents General Information Common SpecificationsTemperature Coefficient Page Model CL25/CL25EC Unique SpecificationsModel CL23A/CL23EC Model CL24Manual Addenda FeaturesModel CL27/CL27EC Safety Precautions Safety InformationUnpacking Your Unit Page Battery Installation Battery INSTALLATION/REPLACEMENTOperation with Rechargeable Battery Memory BackupInitial Turn on Diagnostics and Error MessagesProblem Likely Cause Operating InstructionsKeypad OPR Meter Mode Operation Thermocouples Set-UpHold Meter Mode Operation RTD, Thermistor & Ohms Set-UpData Logging and Recall To Recall Readings To Save ReadingsCalibrator Mode Operation Thermocouples Set-Up To Erase Data Locations Meter ModeCHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER Calibrator Mode Operation RTD, Thermistor & Ohms Set-UpWire Resistance Simulation To Store Outputs Storage and Recall of Calibrator SettingsTo Recall Outputs To Erase Set-Up Locations Calibrator ModeRamp Function CL25 and CL27 Step FunctionAmbient Conditions Service InformationCalibration Procedures Thermocouple Calibration Test Equipment RequiredPage Calibration Adjustments Output Adjust Reading Meter-Mode Zero & Gain Adjustments Sensor StepInput Adjust Reading Calibrator-Mode Zero & Gain Adjustments SensorCalibration Setup, Copper Mode Test Equipment Required Resistance CalibrationResistance Calibration Setup Calibration Adjustments RTD-Thermistor Sensor Unit Step Output Input Adjust Verify ReadingInput Adjust Verify Reading Calibration of Meter-Mode Resistance FunctionsCalibration Verification Equipment Required Model Software RevisionPage Page Alloy Mode Verification Procedure for Thermocouple Functions Page Mechanical Parts Diagram Calibration Procedures Page Page Page Page Page USAWARRANTY/DISCLAIMER Temperature