Omega 27, 25, 26, 24 manual Calibration Adjustments

Page 32

Calibration Adjustments:

NOTE: Do not deviate from the calibration adjustment sequence that follows. This will ensure that adjustments to be stored in EEPROM go to the correct locations.

1.Perform meter-mode adjustments and verifications per Table 3 (use Table 5 for RTD-Thermistor models).

2.Hit OPR key (Flashing 'RTD' annunciator turns off).

3.Install J1 jumper.

4.Set unit to CALIB mode, RTD-100, °F.

5.Switch current source to OPERATE mode.

6Perform calibrator-mode adjustments and verifications per Table 4 (use Table 6 for RTD-Thermistor models).

7.Turn off unit. Remove tape holding battery.

8.Re-install back cover. Resistance-function calibration is complete.

Table 3: Calibration of Meter-Mode Resistance Functions (RTD-100/1000)

 

Sensor

Unit

 

 

Unit

Step

Select

Input

Adjust

Verify

Reading*

 

 

 

 

 

 

1

RTD-100

390.26Ω

P1**

 

1562.0°F±0.5°F

2

RTD-100

100.00Ω

 

32.0±0.1°F

3

RTD-100

18.49Ω

 

–328.0±0.2°F

4

RTD-1000

3,902.6Ω

P1**

 

1562.0°F±0.5°F

5

RTD-1000

1000.0Ω

 

32.0+0.1°F

6

1,000Ω

900.00Ω

P1

 

900.00Ω

7

1,000Ω

1.00Ω

 

1.00±0.10Ω

8

10,000Ω

9,000.0Ω

P1

 

9000.0Ω

9

10,000Ω

10.0Ω

 

10.0±1.0Ω

 

 

 

 

 

 

*exclusive of noise.

**adjust as necessary to obtain steps 2, 5.

30

Image 32
Contents CL23A, 24, 25, 26 Omega.comTM Table of Contents Common Specifications General InformationTemperature Coefficient Page Unique Specifications Model CL23A/CL23ECModel CL24 Model CL25/CL25ECModel CL27/CL27EC FeaturesManual Addenda Unpacking Your Unit Safety InformationSafety Precautions Page Battery INSTALLATION/REPLACEMENT Battery InstallationMemory Backup Operation with Rechargeable BatteryDiagnostics and Error Messages Problem Likely CauseOperating Instructions Initial Turn onKeypad Meter Mode Operation Thermocouples Set-Up OPRMeter Mode Operation RTD, Thermistor & Ohms Set-Up HoldData Logging and Recall To Save Readings To Recall ReadingsTo Erase Data Locations Meter Mode Calibrator Mode Operation Thermocouples Set-UpCalibrator Mode Operation RTD, Thermistor & Ohms Set-Up CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTERWire Resistance Simulation Storage and Recall of Calibrator Settings To Store OutputsTo Erase Set-Up Locations Calibrator Mode To Recall OutputsStep Function Ramp Function CL25 and CL27Service Information Calibration ProceduresThermocouple Calibration Test Equipment Required Ambient ConditionsPage Calibration Adjustments Meter-Mode Zero & Gain Adjustments Sensor Step Input Adjust ReadingCalibrator-Mode Zero & Gain Adjustments Sensor Output Adjust ReadingCalibration Setup, Copper Mode Resistance Calibration Test Equipment RequiredResistance Calibration Setup Calibration Adjustments Output Input Adjust Verify Reading Input Adjust Verify ReadingCalibration of Meter-Mode Resistance Functions RTD-Thermistor Sensor Unit StepCalibration Verification Model Software Revision Equipment RequiredPage Page Alloy Mode Verification Procedure for Thermocouple Functions Page Mechanical Parts Diagram Calibration Procedures Page Page Page Page Page USAWARRANTY/DISCLAIMER Temperature