Calibration Adjustments:
NOTE: Do not deviate from the calibration adjustment sequence that follows. This will ensure that adjustments to be stored in EEPROM go to the correct locations.
1.Perform
2.Hit OPR key (Flashing 'RTD' annunciator turns off).
3.Install J1 jumper.
4.Set unit to CALIB mode,
5.Switch current source to OPERATE mode.
6Perform
7.Turn off unit. Remove tape holding battery.
8.
Table 3: Calibration of
| Sensor | Unit |
|
| Unit |
Step | Select | Input | Adjust | Verify | Reading* |
|
|
|
|
|
|
1 | 390.26Ω | P1** |
| 1562.0°F±0.5°F | |
2 | 100.00Ω |
| √ | 32.0±0.1°F | |
3 | 18.49Ω |
| √ | ||
4 | 3,902.6Ω | P1** |
| 1562.0°F±0.5°F | |
5 | 1000.0Ω |
| √ | 32.0+0.1°F | |
6 | 1,000Ω | 900.00Ω | P1 |
| 900.00Ω |
7 | 1,000Ω | 1.00Ω |
| √ | 1.00±0.10Ω |
8 | 10,000Ω | 9,000.0Ω | P1 |
| 9000.0Ω |
9 | 10,000Ω | 10.0Ω |
| √ | 10.0±1.0Ω |
|
|
|
|
|
|
*exclusive of noise.
**adjust as necessary to obtain steps 2, 5.
30