Calibration Adjustments:
NOTE: Do not deviate from the calibration adjustment sequence that follows. This will ensure that adjustments to be stored in EEPROM go to the correct locations.
1.Set
2.Hit OPR key (RTD annunciator turns off).
3.Set U.U.T. to CALIB, OPR, K, °F.
4.Turn U.U.T. OFF, then ON, to
5.Set
6.Hit OPR key (RTD annunciator turns off).
7.Install J1 jumper.
8.Set U.U.T. to METER, OPR, T, °F
9.Change TC simulator output to Type T, 32°F, ALLOY mode.
10.Reconfigure calibration
11.Calculate E1 as follows: E1 = U.U.T. Reading — 32.0.
12.Set U.U.T. to CALIB, OPR, T, °F mode.
13.Store E1; CHANGE/(E1 value)/ENTER/STO/1/OPR.
14.Go to METER mode, and verify U.U.T. reading of 32.0 ±0.1°F.
15.Reconfigure calibration setup per Figure 12. Leave simulator output at 32°F, Type T, ALLOY mode. Allow
16.Set U.U.T. to CALIB mode, 32°F, Type T.
17.Calculate E2 as follows:
E2 = DMM Reading (in µV)
(For example with DMM reading of +11µV, E2 =
18. Store E2; CHANGE/(E2 value)/ENTER/STO/2/OPR.
19. Install J2 jumper.
20.
21. Turn OFF U.U.T.
22. Thermocouple calibration is complete. Remove calibration cover. Re- install original
Calibration).
25