Omega 24, 27, 25, 26 manual Calibration Adjustments

Page 27

Calibration Adjustments:

NOTE: Do not deviate from the calibration adjustment sequence that follows. This will ensure that adjustments to be stored in EEPROM go to the correct locations.

1.Set meter-mode zero and gain adjustments per Table 1.

2.Hit OPR key (RTD annunciator turns off).

3.Set U.U.T. to CALIB, OPR, K, °F.

4.Turn U.U.T. OFF, then ON, to re-enable RTD & Ω annunciators.

5.Set calibrator-mode zero and gain adjustments per Table 2. Hit CHANGE and ENTER at the completion of each step in Table 2.

6.Hit OPR key (RTD annunciator turns off).

7.Install J1 jumper.

8.Set U.U.T. to METER, OPR, T, °F

9.Change TC simulator output to Type T, 32°F, ALLOY mode.

10.Reconfigure calibration set-up per Figure 12. Allow 2-3 minutes for ther- mal stabilization.

11.Calculate E1 as follows: E1 = U.U.T. Reading — 32.0.

12.Set U.U.T. to CALIB, OPR, T, °F mode.

13.Store E1; CHANGE/(E1 value)/ENTER/STO/1/OPR.

14.Go to METER mode, and verify U.U.T. reading of 32.0 ±0.1°F.

15.Reconfigure calibration setup per Figure 12. Leave simulator output at 32°F, Type T, ALLOY mode. Allow 2-3 minutes for thermal stabilization.

16.Set U.U.T. to CALIB mode, 32°F, Type T.

17.Calculate E2 as follows:

E2 = DMM Reading (in µV)

–22

(For example with DMM reading of +11µV, E2 = –0.5°F)

18. Store E2; CHANGE/(E2 value)/ENTER/STO/2/OPR.

19. Install J2 jumper.

20. Re-set U.U.T. output to 32°F, DMM reading should be 0±2µV.

21. Turn OFF U.U.T.

22. Thermocouple calibration is complete. Remove calibration cover. Re- install original back-cover, unless going on to Part B (Resistance

Calibration).

25

Image 27
Contents CL23A, 24, 25, 26 Omega.comTM Table of Contents General Information Common SpecificationsTemperature Coefficient Page Model CL25/CL25EC Unique SpecificationsModel CL23A/CL23EC Model CL24Features Manual AddendaModel CL27/CL27EC Safety Information Safety PrecautionsUnpacking Your Unit Page Battery Installation Battery INSTALLATION/REPLACEMENTOperation with Rechargeable Battery Memory BackupInitial Turn on Diagnostics and Error MessagesProblem Likely Cause Operating InstructionsKeypad OPR Meter Mode Operation Thermocouples Set-UpHold Meter Mode Operation RTD, Thermistor & Ohms Set-UpData Logging and Recall To Recall Readings To Save ReadingsCalibrator Mode Operation Thermocouples Set-Up To Erase Data Locations Meter ModeCHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER Calibrator Mode Operation RTD, Thermistor & Ohms Set-UpWire Resistance Simulation To Store Outputs Storage and Recall of Calibrator SettingsTo Recall Outputs To Erase Set-Up Locations Calibrator ModeRamp Function CL25 and CL27 Step FunctionAmbient Conditions Service InformationCalibration Procedures Thermocouple Calibration Test Equipment RequiredPage Calibration Adjustments Output Adjust Reading Meter-Mode Zero & Gain Adjustments Sensor StepInput Adjust Reading Calibrator-Mode Zero & Gain Adjustments SensorCalibration Setup, Copper Mode Test Equipment Required Resistance CalibrationResistance Calibration Setup Calibration Adjustments RTD-Thermistor Sensor Unit Step Output Input Adjust Verify ReadingInput Adjust Verify Reading Calibration of Meter-Mode Resistance FunctionsCalibration Verification Equipment Required Model Software RevisionPage Page Alloy Mode Verification Procedure for Thermocouple Functions Page Mechanical Parts Diagram Calibration Procedures Page Page Page Page Page USAWARRANTY/DISCLAIMER Temperature