SERVICE INFORMATION
WARNING
All service information is intended for qualified electronic mainte- nance personnel only.
1. CALIBRATION PROCEDURES
This is a
A.THERMOCOUPLE CALIBRATION Test Equipment Required:
1.Thermocouple simulator (OMEGA CL521, or equivalent) calibrated to
2.Microvolt DMM (OMEGA OM7551, or equivalent)
3.Calibration cables, per Figures 11, 12, 13.
NOTE: The copper constantan cables used in Figures 12 and 13 need to be calibrated. Voltage errors in these cables should be com- pensated at the simulator.
4. Calibration cover (OMEGA
Ambient Conditions:
Units should be calibrated at an ambient temperature of 23°C ±1°C, with rel- ative humidity less than 80%.
Preparation for Calibration:
11. Remove battery from unit under test (U.U.T.). Refer to Battery Installation/Replacement section in the manual.
12. Remove
13. Remove
NOTE: Removal of J1, J2 causes partial loss of previously stored calibration data. Do not remove these jumpers unless recalibra- tion is intended.
23