B. RESISTANCE CALIBRATION
NOTE: Thermocouple and resistance calibrations are indepen- dent. However, potentiometer P1 is used in both thermocouple and resistance calibrations to set EEPROM constants. Follow cal- ibration procedures carefully to ensure that thermocouple and resistance calibrations do not disturb one another.
Test Equipment Required:
1.Resistance Decades (General Resistance
2.Current Source (Fluke 5101B, or equivalent).
3.Digital Multimeter (OMEGA OM7551, or equivalent).
4.Calibration cables per Figure 14.
Ambient Conditions:
Unit should be calibrated at an ambient temperature of 23°C±1°C, with rela- tive humidity less than 80%.
Preparation For Calibration:
1.Turn off unit, and remove
2.Secure battery with a piece of adhesive tape.
NOTE: Momentary loss of battery power during the following cal- ibration will invalidate previous adjustments, both resistance and thermocouple.
3.Hook up
4.Set
5.Set DMM to 2V DC range.
6.Turn on unit.
7.Set unit to METER mode,
8.Turn off unit. Remove jumper J1. Turn on unit. 'RTD' annunciator should be flashing.
NOTES: Flashing 'RTD' indicates that calibration adjustments will be written to memory. Avoid selecting thermocouple functions while in this mode.
NOTES: Should it be necessary to scroll through thermocouple functions, first hit the OPR key (disables flashing 'RTD'), and start again at Step 7 above.
28