Omega 26, 27, 25, 24 manual Alloy Mode Verification Procedure for Thermocouple Functions

Page 38

32.Press the SENSOR SELECT function button. The sensor type will change to type C. A “C” will appear momentarily in the center of the display. “600.0” will then appear in the center of the display and both the “T” and “R” will appear along the top row of the display.

33.Press the CHANGE function button. The “600.0 °F” display will blink, and “NUM” will appear on the display.

34.Press 3 2 0 buttons and then press ENTER function button. “NUM” will dis- appear and “32.0 °F” will appear, not blinking. The voltmeter should read 0.000 mV.

35.Press the SENSOR SELECT function button. The sensor type will change to type D. A “d” will appear momentarily in the center of the display. “32.0” will then appear in the center of the display and both the “T” and “R” will appear along the top row of the display. The voltmeter should read 0.000 mV.

36.Press the SENSOR SELECT function button. The sensor type will change to type K. The “T” and the “R” will disappear from the display and a “K” will appear on the display. The voltmeter should read 0.000 mV.

37.Press the OFF function button. The unit will turn off.

38.Press the ON function button. The unit will power up and return to the cali-

brator mode. “CALIB” will appear on the display along with the “K” enuncia- tor, the “OPR” enunciator, and “32.0 °F” will be displayed. The blinking“Ω ” symbol will have completely disappeared indicating that you have exited the cop- per mode.

39.Remove the Copper Mode Calibrator Verification Cable from both the voltmeter and the CL25.

40.The copper mode verification is now complete. Press the OFF function button to turn the unit off.

ALLOY MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE FUNCTIONS

Equipment Required:

1.OMEGA Model to be verified.

2.Precision temperature simulator / voltage source.

3.Voltmeter accurate to 1µV resolution.

4.Cable assembly.

1.Start your calibration with a fresh battery.

2.Press the ON function button to turn the unit on.

36

Image 38
Contents CL23A, 24, 25, 26 Omega.comTM Table of Contents Common Specifications General InformationTemperature Coefficient Page Model CL24 Unique SpecificationsModel CL23A/CL23EC Model CL25/CL25ECModel CL27/CL27EC FeaturesManual Addenda Unpacking Your Unit Safety InformationSafety Precautions Page Battery INSTALLATION/REPLACEMENT Battery InstallationMemory Backup Operation with Rechargeable BatteryOperating Instructions Diagnostics and Error MessagesProblem Likely Cause Initial Turn onKeypad Meter Mode Operation Thermocouples Set-Up OPRMeter Mode Operation RTD, Thermistor & Ohms Set-Up HoldData Logging and Recall To Save Readings To Recall ReadingsTo Erase Data Locations Meter Mode Calibrator Mode Operation Thermocouples Set-UpCalibrator Mode Operation RTD, Thermistor & Ohms Set-Up CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTERWire Resistance Simulation Storage and Recall of Calibrator Settings To Store OutputsTo Erase Set-Up Locations Calibrator Mode To Recall OutputsStep Function Ramp Function CL25 and CL27Thermocouple Calibration Test Equipment Required Service InformationCalibration Procedures Ambient ConditionsPage Calibration Adjustments Calibrator-Mode Zero & Gain Adjustments Sensor Meter-Mode Zero & Gain Adjustments Sensor StepInput Adjust Reading Output Adjust ReadingCalibration Setup, Copper Mode Resistance Calibration Test Equipment RequiredResistance Calibration Setup Calibration Adjustments Calibration of Meter-Mode Resistance Functions Output Input Adjust Verify ReadingInput Adjust Verify Reading RTD-Thermistor Sensor Unit StepCalibration Verification Model Software Revision Equipment RequiredPage Page Alloy Mode Verification Procedure for Thermocouple Functions Page Mechanical Parts Diagram Calibration Procedures Page Page Page Page Page USAWARRANTY/DISCLAIMER Temperature