NOTE: • With J1 removed, the
NOTE: • Do not deviate from the calibration adjustment sequence that follows. This will ensure that zero adjust- ments for each TC type will be stored in the appropriate memory locations.
18. Apply 32°F (Type K) to U.U.T. Adjust potentiometer P1 for a reading of 32.0°F (exclusive of noise).
19. Apply 2500°F (Type K) to U.U.T. Adjust potentiometer P11 for a reading of 2500.0°F.
NOTE: Adjustment steps 8 & 9 are interactive. Repeat these steps as necessary to get the desired readings.
10.Hit OPR key to store
11.Select Type J thermocouple (KJT key on CL23A, SENSOR SELECT key on CL24).
12.Turn off U.U.T.
13.Replace K
14.Turn on U.U.T. Note that “FUNC” annunciator is on again. Allow
15.Apply 32°F (Type J). Adjust P1 for a reading of 32.0°F (exclusive of noise).
16.Hit OPR key to store
17.Select Type T thermocouple.
18.Turn off U.U.T.
19.Replace J
20.Turn on U.U.T. Allow
21.Apply 32°F (Type T). Adjust P1 for a reading of 32.0°F (exclusive of noise).
22.Hit OPR key to store
NOTE: Steps 23 thru 28 apply to Model CL24 only. For Model CL23A, go to step 29 next.
40