Omega 24, 27, 25, 26 manual Model Software Revision, Equipment Required

Page 35

Model

Software Revision

CL23A

6.7

CL24

7.7

CL25

3.7

CL26

5.6

CL27

4.16

(Software revision is marked on a label attached to integrated-circuit U1 on the mother-board).

"Copper-mode" is a non-compensated mode of operation during which cold-junction compensation of inputs and outputs is disabled. This permits the use of copper-wire interconnections between instruments, while avoid- ing the time and expense of multiple thermocouple-cable hookups. At the conclusion of calibration verification, cold-junction compensation is re- enabled, and operation returns to "alloy-mode".

“Alloy-mode” refers to Thermocouple operations; for example, a Type-K thermocouple is made up of Nickel/Chromium and Nickel/Aluminum alloys. Referring to the OMEGA Calibrator(s) in TC operation, “Alloy-mode” means in effect “other than Copper-mode.” In order to make accurate measure- ments using a thermocouple device one must eliminate erroneous voltage signals. In “Alloy-mode”, the OMEGA calibrator automatically calculates the cold-junction compensation for the selected TC type, based on the value(s) from the ITS-90 table. This feature enables the user to make accurate tem- perature measurements without making manual calculations for error cor- rections.

COPPER MODE VERIFICATION PROCEDURE FOR THERMOCOUPLE FUNCTIONS

Equipment Required:

1.OMEGA Model to be verified.

2.Voltmeter accurate to 1µV resolution.

3.Cable assembly (Fig. 11).

1.Start your calibration with a fresh battery.

2.Press the ON function button to turn the unit on.

3.Select the calibrator mode (“CALIB” on display) by pressing the CALIB/METER function button.

33

Image 35
Contents CL23A, 24, 25, 26 Omega.comTM Table of Contents General Information Common SpecificationsTemperature Coefficient Page Model CL25/CL25EC Unique SpecificationsModel CL23A/CL23EC Model CL24Model CL27/CL27EC FeaturesManual Addenda Unpacking Your Unit Safety InformationSafety Precautions Page Battery Installation Battery INSTALLATION/REPLACEMENTOperation with Rechargeable Battery Memory BackupInitial Turn on Diagnostics and Error MessagesProblem Likely Cause Operating InstructionsKeypad OPR Meter Mode Operation Thermocouples Set-UpHold Meter Mode Operation RTD, Thermistor & Ohms Set-UpData Logging and Recall To Recall Readings To Save ReadingsCalibrator Mode Operation Thermocouples Set-Up To Erase Data Locations Meter ModeCHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER Calibrator Mode Operation RTD, Thermistor & Ohms Set-UpWire Resistance Simulation To Store Outputs Storage and Recall of Calibrator SettingsTo Recall Outputs To Erase Set-Up Locations Calibrator ModeRamp Function CL25 and CL27 Step FunctionAmbient Conditions Service InformationCalibration Procedures Thermocouple Calibration Test Equipment RequiredPage Calibration Adjustments Output Adjust Reading Meter-Mode Zero & Gain Adjustments Sensor StepInput Adjust Reading Calibrator-Mode Zero & Gain Adjustments SensorCalibration Setup, Copper Mode Test Equipment Required Resistance CalibrationResistance Calibration Setup Calibration Adjustments RTD-Thermistor Sensor Unit Step Output Input Adjust Verify ReadingInput Adjust Verify Reading Calibration of Meter-Mode Resistance FunctionsCalibration Verification Equipment Required Model Software RevisionPage Page Alloy Mode Verification Procedure for Thermocouple Functions Page Mechanical Parts Diagram Calibration Procedures Page Page Page Page Page USAWARRANTY/DISCLAIMER Temperature