Instruction Manual

Appendix A Rev. 3.6 July 1998

World Class 3000

PIN

WRENCH

RETAINER

OPTIONAL CERAMIC

DIFFUSION ELEMENT

SETSCREW

 

HUB

CEMENT

CEMENT

PORT

FILLET

 

VEE

DEFLECTOR

21240026

Figure A-17. Removal of Optional Diffusor

and Vee Deflector

PROBE TUBE

(NOT INCLUDED

IN KIT)

CORRUGATED

SEAL

CELL AND

FLANGE

ASSEMBLY

SOCKET HEAD

CAP SCREWS

CALIBRATION GAS

PASSAGE

g.Rub a small amount of anti-seize on both sides of new corrugated seal.

h.Assemble cell and flange assembly, corru- gated seal, and probe tube. Make sure the calibration tube lines up with the calibration gas passage in each component. Apply a small amount of anti-seize compound to screw threads and use screws to secure assembly. Torque to 55 in-lbs (4 Nm).

i.Apply anti-seize compound to threads of cell assembly, hub, and setscrews. Reinstall hub on cell assembly. Using pin spanner wrenches, torque to 10 ft-lbs (14 Nm). If applicable, reinstall vee deflector, orienting apex toward gas flow. Secure with setscrews and anti-seize compound. Torque to 25 in-lbs (2.8 Nm).

j.On systems equipped with an abrasive

shield, install dust seal gaskets, with joints 180o apart.

k.Reinstall probe and gasket on stack flange. If there is an abrasive shield in the stack,

make sure dust seal gaskets are in place as they enter 15o reducing cone.

l.Turn power on to electronics and monitor thermocouple output. It should stabilize at 29.3 ±0.2 mV. Set reference air flow at 2 scfh (56.6 L/hr). After probe stabilizes, calibrate probe per Instruction Bulletin applicable to your electronics package. If new components have been installed, repeat calibration after 24 hours of operation.

 

21240009

A-12 OPTIONAL CERAMIC DIFFUSION

 

 

 

Figure A-18. Cell Replacement Kit

ELEMENT REPLACEMENT

e.

If contact assembly is damaged, replace

a. General

 

contact and thermocouple according to

 

 

paragraph A-13,Replacement of Contact

The diffusion element protects the cell from

 

and Thermocouple Assembly.

particles in process gases. It does not nor-

 

 

mally need to be replaced because the vee

f.

Remove and discard corrugated seal. Clean

deflector protects it from particulate erosion.

 

mating faces of probe tube and retainer.

In severe environments the filter may be

 

Remove burrs and raised surfaces with

broken or subject to excessive erosion. Ex-

 

block of wood and crocus cloth. Clean

amine the diffusion element whenever re-

 

threads on retainer and hub.

moving the probe for any purpose. Replace

 

 

if damaged.

A-18 Appendices

Rosemount Analytical Inc. A Division of Emerson Process Management

Image 52
Contents World Class Essential Instructions Highlights of Changes Effective October, 1995 RevSummary Effective June, 1996 RevEffective May, 1997 Rev PageSummary Effective February, 1998 Rev PageSummaryTable of Contents World ClassList of Illustrations Typical System PackagePreface DefinitionsWorld Class Section Description Component Checklist of Typical System Package ContentsOverview System Configuration FeaturesTypical System Installation Heater Power Supply Probe Head WiringExisting Electronics ProbeSection Installation Either make necessary repairs or installInstallation Oxygen Analyzer ProbeProbe Installation Sheet 1 Probe Installation Sheet 2 Probe Installation Sheet 3 Probe Installation Sheet 4 Probe Installation Sheet 5 Service Required Orienting the Optional Vee DeflectorElectrical Installation of Heater Power Supply Heater Power Supply Installation Electrical ConnectionsSelection JM1 Fuses JM2 World Class Section Setup Electronics SetupModel 218A Electronics Setup G02 G04Eprom Replacement Model TC200 Veritrim Electronics SetupHeater Set Point Adjustment Model 132 Digital Electronics Setup Main PCB Model 132 Eprom ReplacementSection Troubleshooting System TroubleshootingWorld Class Section Return of Material World Class Section Appendices World Class Oxygen Analyzer Probe General Figure A-2. Main Probe ComponentsTable A-1. Specifications for Oxygen Analyzing Equipment.1 Probe Assembly Exterior Cell and Flange AssemblyProbe Tube Assembly Snubber Diffusion AssemblyInner Probe Assembly Cell GeneralCable Assembly Abrasive Shield AssemblyProbe Junction BOX Probe OptionsView a Figure A-8. Ceramic Diffusion/Dust Seal Assembly Ceramic Diffusion AssemblyBypass Probe Options Probe Mounting Jacket OptionsFigure A-13. Bypass Probe Option Sheet 1 Figure A-13. Bypass Probe Option Sheet 2 Extended Temperature By-Pass Arrangements 2400 F 1300 C Group Code DescriptionProbe Troubleshooting Probe FaultsTable A-2. Fault Finding Symptom Check Remedy OverviewWorld Class Figure A-14. Flowchart of Probe Related Problems, #1 Figure A-15. Flowchart of Probe Related Problems, #2 Cell Replacement Probe RecalibrationFigure A-16. Cell Wiring Connection Optional Ceramic Diffusion Element ReplacementGeneral World Class Replacement of Contact Thermocouple Assembly Figure A-19. Probe Junction Box Mechanical ConnectionsContact Heater Screws Not Shown Thermocoupler World Class Figure A-22. Oxygen Analyzer Probe, Cross-Sectional View Figure A-23. High Temperature Corrosive Environment Kit Replacement Parts Figure A-10 4841B03G02 Stainless Steel Diffuser Assembly Appendix B, REV HPS 3000 Heater Power Supply Description Front SideTable B-1. Specifications for Heater Power Supply Theory of OperationHPS 3000 Troubleshooting Overview HPS 3000 TroubleshootingSymptom Figure B-4. HPS Troubleshooting Flowchart, #1Figure B-5. HPS Troubleshooting Flowchart, #2 Figure B-6. HPS Troubleshooting Flowchart, #3 Transformer Replacement Fuse ReplacementMother Board Replacement Daughter Board Replacement Figure B-7. Heater Power Supply, Exploded View Table B-2. Replacement Parts for Heater Power Supply Part Number DescriptionWorld Class Section Index World Class Warranty World Class 3000 Probe Serial No Order No HPS
Related manuals
Manual 88 pages 63.05 Kb Manual 60 pages 24.1 Kb Manual 88 pages 36.41 Kb Manual 32 pages 13.16 Kb Manual 22 pages 3.64 Kb Manual 184 pages 52.9 Kb

3000 specifications

The Emerson 3000 is a cutting-edge control system designed to enhance the efficiency, reliability, and precision of industrial operations. Employed across various sectors such as oil and gas, pharmaceutical, food and beverage, and power generation, the Emerson 3000 has gained recognition for its robustness and versatility.

One of the main features of the Emerson 3000 is its advanced process control capability. With integrated control algorithms, it can optimize complex processes in real-time, resulting in significant improvements in production rates and reduced operational costs. The system's predictive analytics capabilities enable operators to anticipate equipment failures and maintenance needs, allowing for proactive management and minimizing downtime.

The Emerson 3000 features a modular architecture, providing flexibility for scaling and customization. Operators can easily tailor the system to fit specific application needs, whether it requires additional control loops or integration with other systems. This adaptability is particularly beneficial for facilities planning for future expansions or modifications.

Another technology highlight of the Emerson 3000 is its seamless integration with the latest Internet of Things (IoT) advancements. The system is designed to communicate effectively with a variety of smart devices and sensors, harnessing data to create insightful analytics that drive operational excellence. This connectivity empowers businesses to leverage big data for improved decision-making and increased agility.

Additionally, the Emerson 3000 incorporates state-of-the-art cybersecurity measures to safeguard critical data and operations. With built-in security protocols and regular updates, the system protects against emerging cyber threats, ensuring the integrity of the control network.

User experience is also a focal point of the Emerson 3000. The intuitive graphical user interface presents complex data in a user-friendly format, making it easier for operators to monitor system performance and respond to alerts quickly. This ease of use contributes to enhanced safety and operational efficiency.

In summary, the Emerson 3000 represents a fusion of advanced process control, modular design, IoT connectivity, robust cybersecurity, and user-centric interface, making it an ideal choice for industries seeking to enhance their operational performance while adapting to ever-evolving technological landscapes.