Installing Objectives

If you have purchased a standard micro- scope configuration, you will notice the objectives are already installed on the nose- piece and the substage condenser is already installed on the stand. In this case, go to section "Operation". If you purchased your Leica DM750 P by components and not by standard configu-

ration, please continue in this section.

Always rotate the nosepiece by using the knurled nosepiece ring.

1.As you rotate the nosepiece clockwise, attach the objectives by screwing them into the nosepiece holes starting with the lowest magnification and advancing to the highest magnification.

Leica DM750 P

Manual

21

 

 

 

Image 21
Contents Leica DM750 P Manual Chapter overview Contents Safety regulations Safety concept Light bulbs Symbols used in this operating manualImportant information Provide clarityDescription Important NotesContact address EC Declaration of ConformityIntended use Instructions on usePlace of use Non-intended useTransport To any warranty claimsDisposal Integration in third-party productsHealth risks Health risks and dangers of useViral infections of the eye Vidual or detachable eyecupsInformation for the person responsible for the instrument Information for the person responsible for the instrumentCleaning glass surfaces Cleaning coated parts and plastic partsCare instructions General instructionsAccessories Accessories, maintenance and repairMaintenance Repairs and service workElectrical data Fuse ReplacementInput 100-240 V, 50/60 Hz, 5 W 3 W LED EnvironmentLeica DM750 P Introduction Unpacking Check the components against the planned configurationGet Ready Tools used Attaching Analyzer Module or Analyzer/Bertrand Lens ModuleViewing Tube Installing Objectives Substage Condenser Substage CondenserCompletely loosen the two thumbscrews on the condenser mount Get Set Basic setup PowerWork Surface Objective Using the Aperture CondenserSetting up Polarized Light Substage condenser in a swing in/outPolarizer Analyzer into position Focusing FocusingSlot of the right eyetube Viewing Tube AdjustmentEyepiece. Do not raise or lower the stage to focus Slightly, if at allKoehler Set up Koehler Set upObjective Centration ϘϘ Centering keysFocus the polarizing sample Modules Bertrand Lens Operation Image through the eyepiecesYou can now view the conoscopic image through the eyepieces Time Delay Shutoff Changing the status of the time delay shutoffOr Disabled and you will not see any blinking Leica DM750 P Manual Ready! Set! Go Care of the Microscope General General maintenanceHands. There is a handle on the back Is extended This purpose For two compensators in the rearϘϘ It is very important to avoid the excessive Lint-free cloth, lens tissue or cotton swabShould be moistened with solvent, but Around the lensMechanical Maintenance Mechanical MaintenanceDimensions Leica DM750 P Manual Leica DM750 P Manual Leica DM750 P Manual