Leica dm750 p manual Objective Centration, ϘϘ Centering keys, Focus the polarizing sample

Page 32

Objective Centration

Tools used

ϘϘ Centering keys

1.Retrieve the two centering keys from the packaging and insert the centering tools into the centering holes located above the objective in use. Engage the ball end of the centering tools into the set screws, which are hidden in the centering holes.

2.Focus the polarizing sample.

3.Move a noticeable point of the sample to the center of the crosslines.

4.Loosen the stage brake and rotate the stage until the noticeable point of the sample is furthest away from the center of the cross- lines. This may even be outside of the field of view. This indicates the objective is not centered to the stage.

Leica DM750 P

Manual

32

 

 

 

Image 32
Contents Leica DM750 P Manual Chapter overview Contents Safety regulations Safety concept Symbols used in this operating manual Light bulbsImportant information Provide clarityImportant Notes DescriptionContact address EC Declaration of ConformityInstructions on use Intended usePlace of use Non-intended useTo any warranty claims TransportDisposal Integration in third-party productsHealth risks and dangers of use Health risksViral infections of the eye Vidual or detachable eyecupsInformation for the person responsible for the instrument Information for the person responsible for the instrumentCleaning coated parts and plastic parts Cleaning glass surfacesCare instructions General instructionsAccessories, maintenance and repair AccessoriesMaintenance Repairs and service workFuse Replacement Electrical dataInput 100-240 V, 50/60 Hz, 5 W 3 W LED EnvironmentLeica DM750 P Introduction Check the components against the planned configuration UnpackingGet Ready Attaching Analyzer Module or Analyzer/Bertrand Lens Module Tools usedViewing Tube Installing Objectives Substage Condenser Substage CondenserCompletely loosen the two thumbscrews on the condenser mount Get Set Work Surface Basic setupPower Using the Aperture Condenser ObjectivePolarizer Setting up Polarized LightSubstage condenser in a swing in/out Analyzer into position Focusing FocusingViewing Tube Adjustment Slot of the right eyetubeEyepiece. Do not raise or lower the stage to focus Slightly, if at allKoehler Set up Koehler Set upFocus the polarizing sample Objective CentrationϘϘ Centering keys Modules You can now view the conoscopic image through the eyepieces Bertrand Lens OperationImage through the eyepieces Or Disabled and you will not see any blinking Time Delay ShutoffChanging the status of the time delay shutoff Leica DM750 P Manual Ready! Set! Go Care of the Microscope General maintenance GeneralHands. There is a handle on the back Is extended This purpose For two compensators in the rearLint-free cloth, lens tissue or cotton swab ϘϘ It is very important to avoid the excessiveShould be moistened with solvent, but Around the lensMechanical Maintenance Mechanical MaintenanceDimensions Leica DM750 P Manual Leica DM750 P Manual Leica DM750 P Manual