Leica dm750 p manual Fuse Replacement, Electrical data, Input 100-240 V, 50/60 Hz, 5 W 3 W LED

Page 14

Electrical data and ambient conditions

Fuse Replacement

Unplug the instrument before changing any fuses. The Leica DM750 P has two fuses, which are located behind the

power cord receptacle.

Use only fuse type: 5×20 mm, 1 Amp/250V, Quick acting (# 13RFAG30003)

Electrical data

Input: 100-240 V, 50/60 Hz, 5 W (3 W LED)

Environment

 

Temperature for use

+10 °C … +40 °C

 

 

Storage temperature

-20 °C … +52 °C

 

+50 °F … +104 °F

 

 

Manipulation shock

25 mm on 50 mm hard wood

 

 

Transport shock (unpacked)

100 g / 6 ms

 

 

Transport shock (packed)

800 mm free fall

 

 

Transport vibrations (unpacked)

5–200 Hz / 1.5 g

 

Atmospheric pressure "in use" and for storage: 500–1060 mBar

 

 

Humidity "in use" and for storage:

20–90%

 

 

Installation Category II (Overvoltage Category)

 

 

 

Pollution degree 2

 

 

 

Leica DM750 P

Manual

14

 

 

 

Image 14
Contents Leica DM750 P Manual Chapter overview Contents Safety regulations Safety concept Important information Symbols used in this operating manualLight bulbs Provide clarityContact address Important NotesDescription EC Declaration of ConformityPlace of use Instructions on useIntended use Non-intended useDisposal To any warranty claimsTransport Integration in third-party productsViral infections of the eye Health risks and dangers of useHealth risks Vidual or detachable eyecupsInformation for the person responsible for the instrument Information for the person responsible for the instrumentCare instructions Cleaning coated parts and plastic partsCleaning glass surfaces General instructionsMaintenance Accessories, maintenance and repairAccessories Repairs and service workInput 100-240 V, 50/60 Hz, 5 W 3 W LED Fuse ReplacementElectrical data EnvironmentLeica DM750 P Introduction Check the components against the planned configuration UnpackingGet Ready Attaching Analyzer Module or Analyzer/Bertrand Lens Module Tools usedViewing Tube Installing Objectives Substage Condenser Substage CondenserCompletely loosen the two thumbscrews on the condenser mount Get Set Work Surface Basic setupPower Using the Aperture Condenser ObjectivePolarizer Setting up Polarized LightSubstage condenser in a swing in/out Analyzer into position Focusing FocusingEyepiece. Do not raise or lower the stage to focus Viewing Tube AdjustmentSlot of the right eyetube Slightly, if at allKoehler Set up Koehler Set upFocus the polarizing sample Objective CentrationϘϘ Centering keys Modules You can now view the conoscopic image through the eyepieces Bertrand Lens OperationImage through the eyepieces Or Disabled and you will not see any blinking Time Delay ShutoffChanging the status of the time delay shutoff Leica DM750 P Manual Ready! Set! Go Care of the Microscope Hands. There is a handle on the back Is extended General maintenanceGeneral This purpose For two compensators in the rearShould be moistened with solvent, but Lint-free cloth, lens tissue or cotton swabϘϘ It is very important to avoid the excessive Around the lensMechanical Maintenance Mechanical MaintenanceDimensions Leica DM750 P Manual Leica DM750 P Manual Leica DM750 P Manual