Viewing Tube Adjustment

Adjust the eyetubes for your interpupillary distance.

1.Fold or unfold the eyetubes until you see one illuminated circle.

The viewing tubes maintain a constant tube length for all interpupillary settings.

This means that a change of interpupillary distance does not affect parfocality, magnifica- tion, or calibrations that depend on magnifica- tion.

To correctly set the eyepieces, be sure the crosshair eyepiece is seated securely

in the slot of the right eyetube.

2.Focus the crosshair reticle by gripping the knurl and rotating the top of the eyepiece until the reticle is in clear focus.

3.While looking through only the crosshair eyepiece, bring the specimen into focus with the fine focus of the microscope.

It may help to cover your left eye.

4.Now switch to look only through the other eyepiece (focusing eyepiece without crosshair) with the other eye.

This time, focus the specimen by using the focusing capability in the focusing

eyepiece. Do not raise or lower the stage to focus!

5.Grip the knurl on the focusing eyepiece with one hand and rotate the top of the eyepiece with the other hand until the specimen is in focus with this eye and this focusing eyepiece. This corrects for any vision differences between your right eye and left eye.

6.Now switch to a high magnification objec- tive (not an oil objective) and focus the microscope while viewing with both eyes.

The higher magnifications have a shal- lower depth of field. Therefore, after focusing with a high magnification, you will find that when you change to a lower magni- fication, you only have to adjust the fine focus

slightly, if at all.

Leica DM750 P

Manual

30

 

 

 

Image 30
Contents Leica DM750 P Manual Chapter overview Contents Safety regulations Safety concept Important information Symbols used in this operating manualLight bulbs Provide clarityContact address Important NotesDescription EC Declaration of ConformityPlace of use Instructions on useIntended use Non-intended useDisposal To any warranty claimsTransport Integration in third-party productsViral infections of the eye Health risks and dangers of useHealth risks Vidual or detachable eyecupsInformation for the person responsible for the instrument Information for the person responsible for the instrumentCare instructions Cleaning coated parts and plastic partsCleaning glass surfaces General instructionsMaintenance Accessories, maintenance and repairAccessories Repairs and service workInput 100-240 V, 50/60 Hz, 5 W 3 W LED Fuse ReplacementElectrical data EnvironmentLeica DM750 P Introduction Check the components against the planned configuration UnpackingGet Ready Attaching Analyzer Module or Analyzer/Bertrand Lens Module Tools usedViewing Tube Installing Objectives Substage Condenser Substage CondenserCompletely loosen the two thumbscrews on the condenser mount Get Set Basic setup PowerWork Surface Using the Aperture Condenser ObjectiveSetting up Polarized Light Substage condenser in a swing in/outPolarizer Analyzer into position Focusing FocusingEyepiece. Do not raise or lower the stage to focus Viewing Tube AdjustmentSlot of the right eyetube Slightly, if at allKoehler Set up Koehler Set upObjective Centration ϘϘ Centering keysFocus the polarizing sample Modules Bertrand Lens Operation Image through the eyepiecesYou can now view the conoscopic image through the eyepieces Time Delay Shutoff Changing the status of the time delay shutoffOr Disabled and you will not see any blinking Leica DM750 P Manual Ready! Set! Go Care of the Microscope Hands. There is a handle on the back Is extended General maintenanceGeneral This purpose For two compensators in the rearShould be moistened with solvent, but Lint-free cloth, lens tissue or cotton swabϘϘ It is very important to avoid the excessive Around the lensMechanical Maintenance Mechanical MaintenanceDimensions Leica DM750 P Manual Leica DM750 P Manual Leica DM750 P Manual