Leica dm750 p manual Viewing Tube

Page 20

Viewing Tube

1.Slightly unscrew the set screw located on the top of the Analyzer module.

2.Align the shape of the polarizing viewing tube with the shape of the module.

3.Insert the dovetail of the polarizing view- ing tube into the module and tighten the set screw firmly without forcing it. This precisely positions the body onto the opti- cal axis of the microscope.

Install the eyepiece with the crossline reticle into the right eyetube so that the raised feature on the eyepiece aligns and

is seated with the orientation slot in the right eyetube.

4. Insert the eyepieces into the eyetubes.

5.If you wear eyeglasses for microscope view- ing, keep the rubber eyeguards folded down. If you do not wear eyeglasses, you may find it useful to unfold the rubber eyeguards in order to block out ambient room light.

Leica DM750 P

Manual

20

 

 

 

Image 20
Contents Leica DM750 P Manual Chapter overview Contents Safety regulations Safety concept Symbols used in this operating manual Light bulbsImportant information Provide clarityImportant Notes DescriptionContact address EC Declaration of ConformityInstructions on use Intended usePlace of use Non-intended useTo any warranty claims TransportDisposal Integration in third-party productsHealth risks and dangers of use Health risksViral infections of the eye Vidual or detachable eyecupsInformation for the person responsible for the instrument Information for the person responsible for the instrumentCleaning coated parts and plastic parts Cleaning glass surfacesCare instructions General instructionsAccessories, maintenance and repair AccessoriesMaintenance Repairs and service workFuse Replacement Electrical dataInput 100-240 V, 50/60 Hz, 5 W 3 W LED EnvironmentLeica DM750 P Introduction Check the components against the planned configuration UnpackingGet Ready Attaching Analyzer Module or Analyzer/Bertrand Lens Module Tools usedViewing Tube Installing Objectives Substage Condenser Substage CondenserCompletely loosen the two thumbscrews on the condenser mount Get Set Work Surface Basic setupPower Using the Aperture Condenser ObjectivePolarizer Setting up Polarized LightSubstage condenser in a swing in/out Analyzer into position Focusing FocusingViewing Tube Adjustment Slot of the right eyetubeEyepiece. Do not raise or lower the stage to focus Slightly, if at allKoehler Set up Koehler Set upFocus the polarizing sample Objective CentrationϘϘ Centering keys Modules You can now view the conoscopic image through the eyepieces Bertrand Lens OperationImage through the eyepieces Or Disabled and you will not see any blinking Time Delay ShutoffChanging the status of the time delay shutoff Leica DM750 P Manual Ready! Set! Go Care of the Microscope General maintenance GeneralHands. There is a handle on the back Is extended This purpose For two compensators in the rearLint-free cloth, lens tissue or cotton swab ϘϘ It is very important to avoid the excessiveShould be moistened with solvent, but Around the lensMechanical Maintenance Mechanical MaintenanceDimensions Leica DM750 P Manual Leica DM750 P Manual Leica DM750 P Manual