Phonetics 1104 Programming Temperature Calibration, Interrogating Temperature Calibration

Page 72

Sensaphone® Model 1104 User’s Manual

5.6.1Programming Temperature Calibration

1.Press SET.

SET

#

2. Press CALIBRATE.

6

CALIBRATE

3.Press the number (1 to 4) of the selected temperature input to be calibrated.

1

CALL

DELAY

4

MAX CALLS

2

3

VOICE

INTERCALL

REPS

TIME

4.Enter the number required to offset the current temperature reading so a correct reading is obtained.

To program a positive offset number (up to +10 degrees), enter the number on the keypad. The Model 1104 recites the digits as they are pressed.

To program a negative offset number (up to –10 degrees), first press PAUSE. The unit responds with a “beep.” Next, enter the number on the keypad. The unit recites the digits as they are pressed.

5.Press ENTER. The Model 1104 responds:“Enter.”

ENTER

TONE

PM

NOTE

If you find that your calibration offset exceeds more than + 5 or –5 degrees, other complicating factors could be affecting normal operation of the Model 1104. Call Phonetics for technical assistance.

5.6.2Interrogating Temperature Calibration

1.Press WHAT IS.

WHAT

IS

*

2. Press CALIBRATE.

72

Image 72
Contents Model SENSAPHONE, INC Important Safety Instructions Sensaphone Model 1104 User’s Manual FCC Requirements PHONETICS, INCSensaphone Model 1104 User’s Manual Canadian Department of Communications Statement Sensaphone Model 1104 User’s Manual Viii Summary of the Alarm Dial-Out Process Technical SupportUnit ID Number Dial-out Telephone NumbersTone or Pulse Dialing Special Dialing Rings Until AnswerInput Configuration Enable/Disable InputsEstablishing High and Low Temperature Limits Temperature ScaleHigh Sound Alarm Enable/Disable Exit DelayAlarm Acknowledgment Status ReportIntroduction Technical Support Installation Operating Environment+ + + RUN Installing the Telephone Line Alert Inputs Sensor Connected to an Alert Input Two Sensors Wired to Adjacent Inputs ~ ~ Disconnecting the Model 1104 for Storage or Seasonal Use From the unit completely to avoid damage to the circuitrySensaphone Model 1104 User’s Manual Quick Start Local KeypadSet Configuration of Inputs RUNStatus Set the ID NumberSet Dial-Out Telephone Numbers NumberSet Temperature Limits Model 1104 responds Enter low temperature limitModel 1104 responds Enter high temperature limit Quick Start Action-Response Programmable Feature Alert ConditionCommunications Programming Programming the ID NumberHello, this is telephone number Unit ID Number Interrogating the ID Number Programming Dial-out Telephone Numbers Enter numberInterrogating a Dial-out Telephone Number Erasing a Telephone NumberPress ENTER. The Model 1104 will say Enter Tone or Pulse Dialing Special Dialing KeysPause Incorporating a Pause Pound #Asterisk CodeIncorporating a Pound # or Asterisk PressedPress any unassigned number key from 1 to 4 to represent Special Dialing to a Beeper or Pager Now press Pause once to activate a four second delay. This Programming Rings Until Answer Interrogating Rings Until AnswerTAD Enable/Disable Using the TAD FeatureNo TAD In Use Programming the Listen-in Time Interrogating the Listen-in Time Call Delay Programming the Call DelayInterrogating Call Delay Enable/Disable Local Voice Mute Local Voice MuteProgramming Voice Repetitions 10.2 Interrogating Voice RepetitionsModel 1104 will recite the number programmed Programming Intercall Time Interrogating Intercall Time SecondsProgramming Max Calls Interrogating Max Calls Setting the Clock Interrogating for the Current Time Locking the Keypad Unlocking the KeypadUsing the number keys, enter the digits for the code Sensaphone Model 1104 User’s Manual Alarm Programming Programming Input Configuration StatusVerifying Input Configuration Changing Enabled/Disabled Input StatusVerifying Enabled/Disabled Input Status Programming Input Recognition Time Immediately follows the number recited for that inputInterrogating Input Recognition Time Recites the digits as they are pressedPress the corresponding input key 1 to Programming Temperature Limits for a Selected Input Disabling Alarm Response to High or Low Temperature Interrogating High and Low Temperature LimitsTemperature Calibration Programming Temperature Calibration Interrogating Temperature CalibrationObtaining Current Temperature Enabling/Disabling the AC Power Alarm Programming Power Failure Recognition Time Model 1104 will recite the digits as they are pressedInterrogating Power Failure Recognition Time Programming Sound Alarm Sensitivity Using the number keys, enter a value for sound sensitivityProgramming High Sound Alarm Recognition Time Interrogating Sound SensitivityChanging Enabled/Disabled High Sound Alarm Exit Delay Acknowledgment Status Report & Remote Access Local AcknowledgmentTouch-Tone Acknowledgment Callback Acknowledgment Example Status Report, No Alarms Example Status Report, Existing Alarms Example Status Report, Disabled Inputs Battery ConditionRemote Access by Touch-Tone Telephone Battery Condition OK, if over 8.2 VoltsReplace batteries, if below 7.2 Volts Activating Status Report Sensaphone Model 1104 User’s Manual Alarm Detection, Dial-out and Acknowledgment OperationTemperature Valid AlarmCall Delay Alarm MessageAlarm Messages Intercall TimeVoice Repetitions Call ProgressLocal, On-site AcknowledgmentTouch-Tone Fast CodeTAD Enabled Number...the last number it dialedHave a good day Example a Dial-out Telephone Call Appendix a Troubleshooting Problem SolutionCommunications / Dial-out CauseAppendix a Troubleshooting Communications / Dial-out Problem Cause Solution Temperature Monitoring Problem Cause Solution Sound Level Monitoring Problem Cause Solution Other Monitoring Problem Cause Solution101 Appendix a Troubleshooting Be sure to use alkaline batteries-do not Use rechargeable nicad batteriesAppendix B Weekly Testing Procedure 104 Appendix C Accessories Part Number Sensor / SwitchSensaphone Model 1104 User’s Manual 106 Service Department Appendix D Returning the Unit for RepairYear Limited Warranty