Section 2
About the I/O cable
(6)OUT3 (ENABLE output)
This turns ON when the
(7)OUT4 (BUSY output)
This turns ON during sampling with the hold function enabled. It allows you to check whether or not the
(8)Linear output
The linear output outputs a current or voltage in accordance with the measured value.
(9)Linear GND
The linear GND terminal is the 0V terminal for the linear output.
This ground wire must be grounded separately from the other ground wires.
Always ground the linear output terminal even when linear output is not used.
(10) to (13) IN0 to IN3
The following input signal assignments can be selected.
• Signal assignments
| Signal | When [Standard] is selected | When [Bank] is selected |
| (default) | ||
|
|
| |
IN0 |
| External trigger (timing) input | Bank input A |
|
|
|
|
IN1 |
| Reset input | Bank input B |
|
|
|
|
IN2 |
| Unused | Unused |
|
|
|
|
IN3 |
| Zero reset input | Zero reset input |
|
|
|
|
Setting I/O Assignments | |
• Signal functions |
|
|
|
Signal Name | Description |
External trigger (timing) input | This timing input is for signal input from external devices. Use it for hold |
| function timing. |
|
|
Reset input | This resets all executing measurements and outputs. While a reset is |
| being input, judgment output conforms to the |
| ting. If this reset input switches ON while the hold function is used, the |
| state in effect before the hold function was set will be restored. |
|
|
Zero reset input | This is used to execute and clear a zero reset. |
|
|
Bank input A, B | This is used for switching banks. Specify the bank No. in combinations |
| of A and B. |
|
|
For details on the timing charts of external I/O, refer to User’s Manual for the Sensor Controller.
Section 2 INSTALLATION & CONNECTION
User’s Manual