Section 5
Q&A
Q&A
Question | Answer |
When scaling is executed, an error appears and | Scaling cannot be set for one of the following reasons: |
settings cannot be made. | • Scaling has been attempted when the measured value is |
| outside the measuring range. |
| • When |
| between the measured values for the two points is not 1% |
| or more of the rated measuring range. |
|
|
When focusing is executed, an error appears and | Monitor focus settings cannot be made when the distance |
settings cannot be made. | between the two specified points is not 1% or more of the |
| rated measuring range. |
|
|
Section 5 APPENDIX
User’s Manual