Section 5
Error Messages and Countermeasures
Error Messages and Countermeasures
■When [Error] Is Displayed on the Main Display
| Display Details | Cause | Countermeasure | |
LCD screen |
| Overcurrent | One or all of the judgment out- | Cancel the load |
(upper section) |
| puts are | (Recovery is automatic after the | |
|
|
|
| load |
|
|
|
|
|
|
| Calculation data error | One (or all) of the channels | Check the measurement state of |
|
|
| assigned to the operation input | the channels assigned to the oper- |
|
|
| cannot be measured. | ation input, and set all channels to |
|
|
|
| a measurable state. |
|
|
|
|
|
■When all digits on the main display and sub-display blink
| Display Details | Cause | Countermeasure |
|
| |
LCD screen |
| SYSTEM ERROR | The controller is not connected. | Connect the Controller. |
| Section5 |
|
| BANK DATA | ler in error | onds, and then hold the DOWN |
| |
|
| CONNECT |
|
|
|
|
|
| SYSTEM ERROR | Bank data in the Sensor Control- | Hold the UP key down for 3 sec- |
|
|
|
|
|
| key down for 3 seconds. |
|
|
|
| SYSTEM ERROR | Internal error | APPENDIX | ||
|
| Turn the sensor ON again. |
| |||
|
|
|
| The sensor is turned ON again |
|
|
|
|
|
| and restored after the device is |
|
|
|
|
|
| initialized. |
|
|
|
| MAIN COM |
|
|
|
|
|
|
|
|
|
|
|
■ Others
| Display Details | Cause | Countermeasure | |
LCD upper sec- |
| Disp range error | The measurement result | Change the decimal point digit |
tion |
|
| exceeds the number of dis- | setting. |
|
|
| played digits. |
|
|
|
|
|
|
Main Display |
| – – – – – – | • The sensor is standing by | • When hold is set, start sam- |
|
|
| for measurement. | pling and apply the hold |
|
|
| • No channels are assigned | value. |
|
|
| to inputs. (Assignments | • Assign the channels. |
|
|
| are set to “NONE”.) |
|
|
|
|
|
|
User’s Manual