4.4.2 Selecting the units of measurement

Screen display:

unitS

 

Factory default setting:

 

 

MSP400RH-B28

m

 

MSP400RH-N28

ft

i)Note that the factory default units of measurement are dictated by the model type, which may be Imperial (ft) or Metric (m).

The user can reconfigure a Metric unit to be an Imperial or vice-versa by changing the base units (b.unit) of the MSP400 - refer to section 4.7.11

ii)Note that changing the base units will cause the MSP400 to re-start with factory default values in all other parameters.

Changing base units after programming the MSP400 will cause all programmed data to be overwritten with factory default values.

The MSP400 is pre-programmed with selected units of measurement for each of the duties available :

Distance and Level measurement

m , ft, in, none

Flow measurement

l/s,

l/m,

m3/hr,

gal/s,

gal/m,

ft3/m (cfm),

ft3/hr,

none

Contents measurement

l, m3, gal, ft3

To change the units of measurement:

a)Press the green button to display the “unitS” menu entry screen. If metres are the chosen units of measurement, as indicated by the small “”m” below the word “unitS” press the green button to continue commissioning.

b)To change the units of measurement from metres to one of the other options, press the blue button to enter the “unitS” menu. Press the blue button again to allow change of the units of measurement. The current unit of measurement will now be flashing, indicating it may be changed.

c)Press the green button to scroll through the list of available duty options.

Notes.

i)The MSP400 will offer a selection of units of measurement relevant to the chosen duty as shown in the option table above.

The final option in each set is “none”, which appears as a blank screen. This option is available to the user who requires to display in units other than those available in the standard option table. In this case, the user will need to scale the PV according to a suitable scaling factor - refer to section 4.4.6

It is strongly recommended that the user make a note of the scale factor and the resultant units of measurement and retain this on a label within the instrument at all times to avoid later confusion.

ii)When using the green button to scroll through the units of measurement options, allow 2-3 seconds after each button press for the MSP400 to check and display the selection. Pressing the green button continuously simply continues scrolling around the units of measurement option loop.

d)Once the desired unit of measurement is shown (flashing) on the display, press the blue button to select this option. It will now stop flashing.

e)If the chosen units of measurement are incorrect, the edit sequence for the units of measurement can be re-started by pressing the blue button again.

If the chosen units of measurement are correct, press the red button to save the units of measurement to memory and automatically scroll on to the next main menu option: b.rEF.

16

IP2046/IM

Nov 2006

 

Image 16
Contents Mobrey Contents Introduction Pvdf Wetside Installation IP2046/IM IP2046/IM IP2046/IM Do not USE Housing to Tighten Ptfe RL1 Spst IP2046/IM Alarm IP2046/IM IP2046/IM Selecting the units of measurement Screen display 20/4 Ullage Liquid level SPEC. P CYL.F Scale LEUEL@ max Maximum flow entry Screen display IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Loop IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Duty Appendix A1 Main Menu ProgrammingSEt20 Appendix A1Appendix A2 Diagnostics Menu LeuelLoop Appendix A3 Commissioning / Loop Test MenuEngineering Menu Appendix A4Appendix B Default Value Listing Main Menu Parameters ProFDefault Value Listing Engineering Parameters DEAdAppendix C Listing of NON-LINEAR Profiles in the MSP400 SPEC.CAppendix D Hart Communications with the MSP400 Table of MSP400 Parameters IP2046/IM IP2046/IM Instruction & maintenance leaflet