To edit the scale factor

a)Press the blue button to enter the “SCALE” menu and to display the current scale factor in use. If this is correct, press the red button return to the main menu.

If the scale factor is to be edited, press the blue button again to allow editing

The leading digit of the current scale factor will now be flashing, indicating it may be changed.

b)Use the green button to edit the value of the leading digit.

c)Once correct, use the blue button to select the next digit and then the green button to edit as before, or the blue button to move to the next digit. Continue this sequence until all 5 digits are correct.

d)Press the blue button a final time to confirm the new value. None of the digits should now be flashing.

If the new scale factor value is incorrect, the edit sequence for the scale factor can be re-started by pressing the blue button again.

If the new scale factor is correct, press the red button to save the new value to memory and automatically scroll on to the next main menu option: “4”

However, if the data already entered allows the MSP400 to calculate the maximum flow, the 4mA and 20mA points are automatically set to 4mA at zero flow and 20mA at maximum flow.

The next menu item offered in this case is “d”. Refer to section 4.4.11

4.4.7 Maximum level entry

Screen display

LEUEL@ max

 

Factory default value

1.000

This selection is offered only if the user has chosen a duty of Flow which requires entry of the level at which the maximum flow occurs, or is shown when manually navigating the main menu - ignore if duty chosen is Level, Distance or Contents.

To enter the value for the level at which the maximum flow occurs:

a)Press the blue button to enter the “LEUEL” @ max” menu and press the blue button again to allow editing.

The leading digit of the level value will now be flashing, indicating it may be changed.

b)Use the green button to edit the value of the leading digit.

Once correct, use the blue button to select the next digit and then the green button to edit as before, or the blue button to move to the next digit. Continue this sequence until all 5 digits are correct.

c)Press the blue button a final time to confirm the new value. None of the digits should now be flashing.

d)If the new level value is incorrect, the edit sequence for the maximum level can be re-started by pressing the blue button again.

e)If the level value entered is correct, press the red button to save the value to memory and automatically scroll on to the next main menu option: “Flo @ max”.

21

IP2046/IM

Nov 2006

 

Image 21
Contents Mobrey Contents Introduction Pvdf Wetside Installation IP2046/IM IP2046/IM IP2046/IM Do not USE Housing to Tighten Ptfe RL1 Spst IP2046/IM Alarm IP2046/IM IP2046/IM Selecting the units of measurement Screen display 20/4 Ullage Liquid level SPEC. P CYL.F Scale LEUEL@ max Maximum flow entry Screen display IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Loop IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Appendix A1 Main Menu Programming DutyAppendix A1 SEt20Leuel Appendix A2 Diagnostics MenuAppendix A3 Commissioning / Loop Test Menu LoopAppendix A4 Engineering MenuProF Appendix B Default Value Listing Main Menu ParametersDEAd Default Value Listing Engineering ParametersSPEC.C Appendix C Listing of NON-LINEAR Profiles in the MSP400Appendix D Hart Communications with the MSP400 Table of MSP400 Parameters IP2046/IM IP2046/IM Instruction & maintenance leaflet