Emerson MSP400RH, IP2046IM manual IP2046/IM

Page 26

4.4.15 Setting the 4 and 20mA levels using actual liquid levels in the tank

 

 

Screen display

(SEt 4 & SEt 20)

Factory default value

Hold

If you have already programmed the 4 and 20mA levels as above, you do not need to enter this menu. All the programming is now complete and you should press the red button to exit the programming menu and return to the main PV display.

If, however, you wish to set the 4 or 20mA level by ranging the instrument to a known target – perhaps the level in the vessel at this time – then press the blue button to enter this menu.

To set the 4mA level

a)With the MSP400 aimed at a target a distance away equivalent to the 4mA level, press the blue button to display the value for the 4mA setting. Press the blue button to allow setting, and the display will alternately flash “4” with the current level reading.

b)Press blue button to confirm the correct value for the 4mA level then press red button to save the new value for the 4mA setting.

c)With the MSP400 aimed at a target a distance away equivalent to the 20mA level, press the blue button to display the value for the 20mA setting. Press the blue button to allow setting, and the display will alternately flash “20” with the current level reading.

d)Press blue button to confirm the correct value for the 4mA level then press red button to save the new value for the 20mA setting.

Programing of the MSP400 is now complete.

Check the main display to ensure the duty, units and PV are correct, and that relays are on or off according to the set points programmed.

The cover may now be replaced, refer back to section 3.3.2

26

IP2046/IM

Nov 2006

 

Image 26
Contents Mobrey Contents Introduction Pvdf Wetside Installation IP2046/IM IP2046/IM IP2046/IM Do not USE Housing to Tighten Ptfe RL1 Spst IP2046/IM Alarm IP2046/IM IP2046/IM Selecting the units of measurement Screen display 20/4 Ullage Liquid level SPEC. P CYL.F Scale LEUEL@ max Maximum flow entry Screen display IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Loop IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Duty Appendix A1 Main Menu ProgrammingSEt20 Appendix A1Appendix A2 Diagnostics Menu LeuelLoop Appendix A3 Commissioning / Loop Test MenuEngineering Menu Appendix A4Appendix B Default Value Listing Main Menu Parameters ProFDefault Value Listing Engineering Parameters DEAdAppendix C Listing of NON-LINEAR Profiles in the MSP400 SPEC.CAppendix D Hart Communications with the MSP400 Table of MSP400 Parameters IP2046/IM IP2046/IM Instruction & maintenance leaflet