Emerson IP2046IM, MSP400RH manual IP2046/IM

Page 9

Transmitter

front face

Flow

Channel invert

Primary device (eg.

 

flume, weir) invert

In addition to the above, when setting the bottom reference on a ‘V’ notch weir it is important that the true invert of the weir is taken and not the meniscus liquid level, which may be 3 to 4mm (1/8”) above the true invert.

Transmitter bottom

reference

Meniscus

True invert

The liquid surface at the point of measurement must have a stable, smooth surface and uniform approach velocity. It must not be affected by baffles, foam, hydraulic jumps or any other object likely to cause flow disruption.

The primary element should be free from any situation where it is likely to ‘drown’ (refer to relevant standard for further information)

The MSP400RH transmitter has integral temperature compensation and must be protected at all times from direct sunlight and any radiated heat.

For maximum accuracy and stability of level measurement reading the transmitter should always be shrouded to prevent the incidence of direct sunlight.

If the flow structure permits, mount the transmitter within the flow channel or chamber. Alternatively, the MSP400RH transmitter can accept an input from an external temperature sensor. See section 3.3.1.

If you are in doubt about any aspect of transmitter installation, contact Mobrey (Service Division) who will be pleased to advise.

9

IP2046/IM

Nov 2006

 

Image 9
Contents Mobrey Contents Introduction Pvdf Wetside Installation IP2046/IM IP2046/IM IP2046/IM Do not USE Housing to Tighten Ptfe RL1 Spst IP2046/IM Alarm IP2046/IM IP2046/IM Selecting the units of measurement Screen display 20/4 Ullage Liquid level SPEC. P CYL.F Scale LEUEL@ max Maximum flow entry Screen display IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Loop IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Appendix A1 Main Menu Programming DutyAppendix A1 SEt20Leuel Appendix A2 Diagnostics MenuAppendix A3 Commissioning / Loop Test Menu LoopAppendix A4 Engineering MenuProF Appendix B Default Value Listing Main Menu ParametersDEAd Default Value Listing Engineering ParametersSPEC.C Appendix C Listing of NON-LINEAR Profiles in the MSP400Appendix D Hart Communications with the MSP400 Table of MSP400 Parameters IP2046/IM IP2046/IM Instruction & maintenance leaflet