Emerson MSP400RH, IP2046IM manual Pvdf

Page 4

2.2 Pressure Equipment Directive

The MSP400RH transmitter does not fall within the PED definition as enclosing a pressurised fluid, so is therefore outside the scope of the Directive.

Accordingly, the Declaration of Conformity does not list the Pressure Equipment Directive.

2.3 Specifications

 

Materials of construction:

 

Transducer material

PVDF

Body & cover Material

Glass filled nylon

Cable gland

Nylon with Nitrile cable seal

Cover seal

Silicone rubber

Cover screws

316 Stainless Steel

Transducer bodyseal

EPDM

Electrical

 

Supply voltage

12-40V dc

Output

4-20mA (3.8 - 20.5mA linear)

 

Relays (2) : SPST rated 1A at 24Vdc

Communications

HART Digital communications (Rev. 5)

Earthing

None required

Recommended Cable

Two core each 0.22 mm2 min, shielded

Max. cable length

3000m

Cable resistance

0.1 Ohm per metre length max.

Cable gland

Suitable for cable sizes 4mm - 8mm diameter

Operating

 

Range

0.45 to 11m

Temperature

 

Ambient

-40°C to +70°C

Wetside

-30°C to +70°C

Pressure

-0.25 bar to 3.0 bar

Weatherproofing

IP66/IP67

Certification

 

The MSP400 range of products are Factory Mutual (Fm) certified as Process Control Equipment meeting basic electrical, mechanical and fire protection requirements (IP66/IP67).

4

IP2046/IM

Nov 2006

 

Image 4
Contents Mobrey Contents Introduction Pvdf Wetside Installation IP2046/IM IP2046/IM IP2046/IM Do not USE Housing to Tighten Ptfe RL1 Spst IP2046/IM Alarm IP2046/IM IP2046/IM Selecting the units of measurement Screen display 20/4 Ullage Liquid level SPEC. P CYL.F Scale LEUEL@ max Maximum flow entry Screen display IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Loop IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Duty Appendix A1 Main Menu ProgrammingSEt20 Appendix A1Appendix A2 Diagnostics Menu LeuelLoop Appendix A3 Commissioning / Loop Test MenuEngineering Menu Appendix A4Appendix B Default Value Listing Main Menu Parameters ProFDefault Value Listing Engineering Parameters DEAdAppendix C Listing of NON-LINEAR Profiles in the MSP400 SPEC.CAppendix D Hart Communications with the MSP400 Table of MSP400 Parameters IP2046/IM IP2046/IM Instruction & maintenance leaflet