4.4.8 Maximum flow entry

Screen display

Flo @ max

 

Factory default value

1.000

This selection is offered only if the user has chosen a duty of Flow which requires entry of the maximum flow capability of the chosen structure (not the maximum flow expected in the application), or is shown when manually navigating the main menu - ignore if duty chosen is Level, Distance or Contents.

To enter the value for maximum flow:

a)Press the blue button to enter the “Flo @ max” menu and press the blue button again to allow editing.

The leading digit of the maximum flow will now be flashing, indicating it may be changed.

b)Use the green button to edit the value of the leading digit.

Once correct, use the blue button to select the next digit and then the green button to edit as before, or the blue button to move to the next digit. Continue this sequence until all 5 digits are correct.

c)Press the blue button a final time to confirm the new value. None of the digits should now be flashing.

d)If the new maximum flow value is incorrect, the edit sequence for the maximum flow can be re-started by pressing the blue button again.

e)If the maximum flow value entered is correct, press the red button to save the value to memory and automatically scroll on to the next main menu option: “4”.

However, with a duty of Flow chosen if the data already entered allows the MSP400 to calculate the maximum flow, the 4mA and 20mA points are automatically set to 4mA at 0 flow and 20mA at maximum flow.

The next menu item offered in this case is “d”.

Refer to section 4.4.12.

4.4.9 Maximum contents entry

Screen display

Cont @ max

 

Factory default value

1.000

This selection is offered only if the user has chosen a duty of Contents which requires entry of the maximum contents of the vessel, or is shown when manually navigating the main menu - ignore if duty chosen is Level, Distance or Flow.

To enter the value for maximum contents:

a)Press the blue button to enter the “Cont @ max” menu and press the blue button again to allow editing.

The leading digit of the maximum contents will now be flashing, indicating it may be changed.

b)Use the green button to edit the value of the leading digit.

Once correct, use the blue button to select the next digit and then the green button to edit as before, or the blue button to move to the next digit. Continue this sequence until all 5 digits are correct.

c)Press the blue button a final time to confirm the new value. None of the digits should now be flashing.

d)If the new maximum contents value is incorrect, the edit sequence for the maximum contents can be re-started by pressing the blue button again.

e)If the maximum contents value entered is correct, press the red button to save the value to memory and automatically scroll on to the next main menu option: “4”.

22

IP2046/IM

Nov 2006

 

Image 22
Contents Mobrey Contents Introduction Pvdf Wetside Installation IP2046/IM IP2046/IM IP2046/IM Do not USE Housing to Tighten Ptfe RL1 Spst IP2046/IM Alarm IP2046/IM IP2046/IM Selecting the units of measurement Screen display 20/4 Ullage Liquid level SPEC. P CYL.F Scale LEUEL@ max Maximum flow entry Screen display IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Loop IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Duty Appendix A1 Main Menu ProgrammingSEt20 Appendix A1Appendix A2 Diagnostics Menu LeuelLoop Appendix A3 Commissioning / Loop Test MenuEngineering Menu Appendix A4Appendix B Default Value Listing Main Menu Parameters ProFDefault Value Listing Engineering Parameters DEAdAppendix C Listing of NON-LINEAR Profiles in the MSP400 SPEC.CAppendix D Hart Communications with the MSP400 Table of MSP400 Parameters IP2046/IM IP2046/IM Instruction & maintenance leaflet