Emerson MSP400RH, IP2046IM manual IP2046/IM

Page 32

4.7.5 Setting the Pulse repetition frequency

Screen display

“Prf”

 

Factory default value:

1.0

The rate of pulses transmitted by the MSP400 is set at a factory default value of once per second.

The MSP400 may be set to transmit faster or more slowly at selected repetition rates between 0.5 and 2.0 times per second.

The pulse repetition frequency may be changed to overcome cross talk problems if more than one ultrasonic transmitter is mounted in the same tank.

To change the pulse repetition frequency:

a)Press the blue button to enter the pulse repetition frequency menu and to display the current value in use. If this is correct, press the red button return to the main menu.

If the pulse repetition frequency is to be changed, press the blue button again to allow editing The pulse repetition frequency value currently in use will now be shown flashing on the display.

b)Use the green button to scroll through the options available (0.5 to 2.0).

c)Use the blue button to select the chosen pulse repetition frequency.

If the new pulse repetition frequency is incorrect, the edit sequence can be re-started by pressing the blue button again.

If the pulse repetition frequency is correct, press the red button to save the new pulse repetition frequency to memory and automatically scroll on to the next main menu option: “Stir”

4.7.6 Setting a valid echo count

Screen display

“Stir”

 

Factory default value:

4

This parameter is normally used in vessels with a stirrer or agitator, particularly if it is slow moving, it is possible for the MSP400 to detect uncovered blades and treat them as a valid echo, thus calculating an incorrect level reading.

The MSP400 may be set to one of a list of pre-selected times in the range 1 to 100

Lowering = faster response

To change the valid echo count :

a)Press the blue button to enter the stirrer menu and to display the current value in use. If this is correct, press the red button return to the main menu.

If the valid echo count is to be changed, press the blue button again to allow editing

The valid echo count currently in use will now be shown flashing on the display.

b)Use the green button to scroll through the options available (1 to 100).

c)Use the blue button to select the chosen valid echo count.

If the valid echo count is incorrect, the edit sequence can be re-started by pressing the blue button again.

If the valid echo count is correct, press the red button to save the valid echo count to memory and automatically scroll on to the next main menu option: “SPi”

32

IP2046/IM

Nov 2006

 

Image 32
Contents Mobrey Contents Introduction Pvdf Wetside Installation IP2046/IM IP2046/IM IP2046/IM Do not USE Housing to Tighten Ptfe RL1 Spst IP2046/IM Alarm IP2046/IM IP2046/IM Selecting the units of measurement Screen display 20/4 Ullage Liquid level SPEC. P CYL.F Scale LEUEL@ max Maximum flow entry Screen display IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Loop IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM IP2046/IM Duty Appendix A1 Main Menu ProgrammingSEt20 Appendix A1Appendix A2 Diagnostics Menu LeuelLoop Appendix A3 Commissioning / Loop Test MenuEngineering Menu Appendix A4Appendix B Default Value Listing Main Menu Parameters ProFDefault Value Listing Engineering Parameters DEAdAppendix C Listing of NON-LINEAR Profiles in the MSP400 SPEC.CAppendix D Hart Communications with the MSP400 Table of MSP400 Parameters IP2046/IM IP2046/IM Instruction & maintenance leaflet