Page 10
Required Equipment
Required equipment is listed in Table B. If the recommended models are not available, equipment with equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Table B. Required Equipment
Equipment | Required Characteristics | Recommended Model |
|
|
|
Calibrator | AC Voltage Range: 0 ~ 750V AC | Fluke 5500 or Wavetek |
| Accuracy: ±0.07% (Basic) | 9100 Calibrator or |
| Frequency Range: 40 ~ 1KHz | equipment |
|
| |
| Accuracy: ±2% |
|
| DC Voltage Range: 0 ~ 1000V DC |
|
| Accuracy: ±0.006% (Basic) |
|
| Current Range: 0 ~ 10A |
|
| Accuracy: AC (40Hz to 1KHz): ±0.08% (Basic) |
|
| DC: ±0.02% (Basic) |
|
| Frequency Source: 1.00Hz ~ 10MHz |
|
| Accuracy: ±0.001% |
|
| Amplitude: 0.5V |
|
| Accuracy: ±5% |
|
| Resistance Range: 1Ω ~ 100MΩ |
|
| Accuracy: ±0.03% (Basic) |
|
| Capacitance Range: 1pF ~ 10mF |
|
| Accuracy: ±0.10% (Basic) |
|
|
|
|
Form Number | Rev 2 November 2007 |