IDEAL INDUSTRIES 61-340 Required Equipment, AC Voltage Range 0 ~ 750V AC, Current Range 0 ~ 10A

Page 12

Page 10

Required Equipment

Required equipment is listed in Table B. If the recommended models are not available, equipment with equivalent specifications may be used.

Repairs or servicing should be performed only by qualified personnel.

Table B. Required Equipment

Equipment

Required Characteristics

Recommended Model

 

 

 

Calibrator

AC Voltage Range: 0 ~ 750V AC

Fluke 5500 or Wavetek

 

Accuracy: ±0.07% (Basic)

9100 Calibrator or

 

Frequency Range: 40 ~ 1KHz

equipment

 

 

 

Accuracy: ±2%

 

 

DC Voltage Range: 0 ~ 1000V DC

 

 

Accuracy: ±0.006% (Basic)

 

 

Current Range: 0 ~ 10A

 

 

Accuracy: AC (40Hz to 1KHz): ±0.08% (Basic)

 

 

DC: ±0.02% (Basic)

 

 

Frequency Source: 1.00Hz ~ 10MHz

 

 

Accuracy: ±0.001%

 

 

Amplitude: 0.5V p-p ~ 5.0V p-p (square wave)

 

 

Accuracy: ±5%

 

 

Resistance Range: 1Ω ~ 100MΩ

 

 

Accuracy: ±0.03% (Basic)

 

 

Capacitance Range: 1pF ~ 10mF

 

 

Accuracy: ±0.10% (Basic)

 

 

 

 

Form Number TM61340-2

Rev 2 November 2007

Image 12
Contents Ideal INDUSTRIES, INC. Technical Manual Table of Contents Table A. The Symbols IntroductionSafety Specifications General specifications Characteristics DescriptionRange Resolution Accuracy Over voltage Protection Range Resolution Accuracy Input ProtectionInput Impedance 10MΩ Frequency response 40~400HzRange Resolution Accuracy Range Resolution Accuracy Max. Test Max. Open Current Range Resolution Sensitivity Accuracy Overload ProtectionRange 8 14 REL 15 Hz/DUTY16 MAX/MIN Physical and Environmental Characteristics Description Environmental characteristics DescriptionCertifications and compliances Required Equipment AC Voltage Range 0 ~ 750V ACFrequency Range 40 ~ 1KHz Accuracy ±2% DC Voltage Range 0 ~ 1000V DCReplacing the Battery Opening the Meter CaseFuse Replacement Replacing FusesTesting the Display LCD Graphics 61-340Testing the Voltage Function for 61-340 Step Range Input ReadingTesting the DC microamperes Function for 61-340 Testing the DC milli amperes Function for 61-340Step Range Input Frequency Reading Step Range SourceTesting the DC ampere Function for 61-340 Testing the AC microampere Function for 61-340Step Range Source Frequency Reading Testing the AC milliampere function for 61-340and Testing the AC a function for 61-340Step Source Frequency Reading Testing the Resistance Function for 61-340 Testing the Capacitance Function for 61-340Step Range Source Reading Checking the Diode Test Function for 61-340 Checking the Continuity Function for 61-340Testing the Frequency Function for 61-340 Testing the Duty Cycle Function for 61-340Testing the Temperature Function for 61-340 DC V Calibration Adjust VR1 AC V Calibration Adjust VR2 ModelTemp Calibration Adjust VR3