Page 6
(5) Resistance (for 
Range | Resolution  | Accuracy  | Over voltage  | |
protection  | ||||
  | 
  | 
  | ||
400.0Ω *1  | 0.1Ω  | ±(1.2% reading + 5 digits)  | 
  | |
4.000KΩ  | 1Ω  | 
  | 
  | |
40.00KΩ  | 10Ω  | ±(1.0% reading + 2 digits)  | 600Vrms  | |
400.0KΩ  | 100Ω  | 
  | ||
  | 
  | |||
4.000MΩ  | 1KΩ  | ±(1.2% reading + 2 digits)  | 
  | |
40.00MΩ *2  | 10KΩ  | ±(2.0% reading + 5 digits)  | 
  | 
Open circuit Voltage: 0.44V approx. *1 < 5 digit of reading rolling.
*2 < 2% of reading rolling.
(6) Diode Check and Continuity (for 
Range | Resolution | Accuracy | Max. Test | Max. Open | |
Current | Circuit Voltage  | ||||
  | 
  | 
  | |||
  | 1mV  | Not specified *  | <1mA, approx.  | 2.5V, approx.  | 
Overload Protection: Not specified
Continuity: 
(7) Capacitance (for 
Range | Resolution  | Accuracy  | Over voltage  | |
Protection  | ||||
  | 
  | 
  | ||
40.00nF  | 10pF  | REL:±(3% reading + 10 digits)  | 
  | |
400.0nF  | 100pF  | 
  | 
  | |
4.000μF  | 1nF  | ±(3% reading + 5 digits)  | 
  | |
40.00μF  | 10nF  | 
  | 
  | |
400.0μF  | 100nF  | ±(20% reading + 5 digits) *2  | 
  | |
4000μF *1  | 10μF  | 
  | ||
  | 
  | 
*1 In this range the reading maybe rolling within specification. *2 Specify reading < half full scale of range.
Note: The meter selects the proper range automatically. Each measurement takes about one second per range, Readings >40.00μF will take ≈ 8 seconds or greater
Form Number   | Rev 2 November 2007  |