6.Using a small flat-tipped screwdriver, adjust VRtr until the display reads 2.995 to 3.005Volts

7.Disconnect the AC calibrator from the meter.

(C)TEMP Calibration (Adjust VR3)

1.Set the rotary switch to the " ºF/ ºC " position.

2.Set the output of the calibrator to 300 ºC

3.Connect the calibrator temperature output to the and COM input terminals

4.Using a small flat-tipped screwdriver to adjust VR3 for a 300 ºC +/- 1 ºC (299 to 301) display

Form Number TM61340-2

Rev 2 November 2007

Image 25
Contents Ideal INDUSTRIES, INC. Technical Manual Table of Contents Introduction Table A. The SymbolsSafety General specifications Characteristics Description SpecificationsRange Resolution Accuracy Input Protection Range Resolution Accuracy Over voltage ProtectionInput Impedance 10MΩ Frequency response 40~400HzRange Resolution Accuracy Range Resolution Accuracy Max. Test Max. Open Current Overload Protection Range Resolution Sensitivity AccuracyRange 15 Hz/DUTY 8 14 REL16 MAX/MIN Environmental characteristics Description Physical and Environmental Characteristics DescriptionCertifications and compliances AC Voltage Range 0 ~ 750V AC Required EquipmentFrequency Range 40 ~ 1KHz Accuracy ±2% DC Voltage Range 0 ~ 1000V DCOpening the Meter Case Replacing the BatteryReplacing Fuses Fuse ReplacementLCD Graphics 61-340 Testing the DisplayStep Range Input Reading Testing the Voltage Function for 61-340Testing the DC milli amperes Function for 61-340 Testing the DC microamperes Function for 61-340Step Range Input Frequency Reading Step Range SourceTesting the AC microampere Function for 61-340 Testing the DC ampere Function for 61-340Step Range Source Frequency Reading Testing the AC a function for 61-340 Testing the AC milliampere function for 61-340andStep Source Frequency Reading Testing the Capacitance Function for 61-340 Testing the Resistance Function for 61-340Step Range Source Reading Checking the Continuity Function for 61-340 Checking the Diode Test Function for 61-340Testing the Duty Cycle Function for 61-340 Testing the Frequency Function for 61-340Testing the Temperature Function for 61-340 AC V Calibration Adjust VR2 Model DC V Calibration Adjust VR1Temp Calibration Adjust VR3