IDEAL INDUSTRIES 61-340 technical manual Step Range Input Frequency Reading, Step Range Source

Page 17

 

 

 

 

 

 

Page 15

Table 2 AC Voltage Test:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Step

Range

Input

Frequency

Reading 61-340

Reading 61-342

 

 

 

 

 

 

 

 

 

 

1

400mV

300mV

60

295.6 to304.4

295.6 to304.4

 

 

 

 

 

 

 

 

 

 

2

400mV

300mV

400

295.6 to304.4

295.6 to304.4

 

 

 

 

 

 

 

 

 

 

3

4V

3.000V

60Hz

2.965 to 3.035

2.965 to 3.035

 

 

 

 

 

 

 

 

 

 

4

4V

3.000V

400Hz

2.965 to 3.035

2.965 to 3.035

 

 

 

 

 

 

 

 

 

 

5

40V

30.00V

60Hz

29.65 to 30.35

29.65 to 30.35

 

 

 

 

 

 

 

 

 

 

6

40V

30.00V

400Hz

29.65 to 30.35

29.65 to 30.35

 

 

 

 

 

 

 

 

 

 

7

400V

300.0V

60Hz

296.5 to 303.5

296.5 to 303.5

 

 

 

 

 

 

 

 

 

 

8

400V

300.0V

400Hz

296.5 to 303.5

296.5 to 303.5

 

 

 

 

 

 

 

 

 

 

9

750V

750V

60Hz

736 to 764

736 to 764

 

 

 

 

 

 

 

 

 

 

10

750V

750V

400Hz

736 to 764

736 to 764

 

 

 

 

 

 

 

 

 

Testing the DC microamperes Function (for 61-340 and 61-342)

To verify the accuracy of AC and DC current measurement functions, do the following:

1.Connect the calibrator to the and COM inputs on the meter.

2.Turn the rotary switch to

a. Press the SELECT button to select the DC function

3.Apply the inputs for steps 1-3 in Table 3a.

4.For each input, compare the reading on the meter display to the reading for your meter in Table 3a

5.If the display reading falls outside of the range shown in Table 3a, the meter does not meet specification.

Table 3a DC microamperes Test:

Step

Range

Source

1

400µA

4.0µA

 

 

 

2

400µA

300.0µA

 

 

 

3

4000µA

3000µA

 

 

 

Reading

3.5to 4.5

297.4to 302.6

2974 to 3026

Testing the DC milli amperes Function (for 61-340 and 61-342)

1.Connect the calibrator to the and COM inputs on the meter.

2.Turn the rotary switch to .

a. Press the SEL button to select the DC function

3.Apply the inputs for steps 1-4 in Table 3b.

4.For each input, compare the reading on the meter display to the reading for your meter in Table 3b

5.If the display reading falls outside of the range shown in Table 3b, the meter does not meet specification.

Form Number TM61340-2

Rev 2 November 2007

Image 17
Contents Ideal INDUSTRIES, INC. Technical Manual Table of Contents Introduction Table A. The SymbolsSafety General specifications Characteristics Description SpecificationsRange Resolution Accuracy Input Protection Range Resolution Accuracy Over voltage ProtectionInput Impedance 10MΩ Frequency response 40~400HzRange Resolution Accuracy Range Resolution Accuracy Max. Test Max. Open Current Range Range Resolution Sensitivity AccuracyOverload Protection 16 MAX/MIN 8 14 REL15 Hz/DUTY Certifications and compliances Physical and Environmental Characteristics DescriptionEnvironmental characteristics Description AC Voltage Range 0 ~ 750V AC Required EquipmentFrequency Range 40 ~ 1KHz Accuracy ±2% DC Voltage Range 0 ~ 1000V DCOpening the Meter Case Replacing the BatteryReplacing Fuses Fuse ReplacementLCD Graphics 61-340 Testing the DisplayStep Range Input Reading Testing the Voltage Function for 61-340Testing the DC milli amperes Function for 61-340 Testing the DC microamperes Function for 61-340Step Range Input Frequency Reading Step Range SourceStep Range Source Frequency Reading Testing the DC ampere Function for 61-340Testing the AC microampere Function for 61-340 Step Source Frequency Reading Testing the AC milliampere function for 61-340andTesting the AC a function for 61-340 Step Range Source Reading Testing the Resistance Function for 61-340Testing the Capacitance Function for 61-340 Checking the Continuity Function for 61-340 Checking the Diode Test Function for 61-340Testing the Duty Cycle Function for 61-340 Testing the Frequency Function for 61-340Testing the Temperature Function for 61-340 AC V Calibration Adjust VR2 Model DC V Calibration Adjust VR1Temp Calibration Adjust VR3