Page 22

Page 20

Testing the Frequency Function (for 61-340 / 61-342)

To verify the accuracy of the meter’s frequency function, do the following:

1. Connect the calibrator to the

and COM inputs on the meter.

2.Note: The accuracy of the calibrator’s frequency function must be appropriate for the specified accuracy of the meter.

3.Set the rotary switch to .

4.Set the calibrator or function generator for the square wave voltage and frequency for steps 1-6 of Table 7a.

5.Compare the reading on the meter display with the display reading shown in Table 7a.

6.If the display reading falls outside of the range shown in Table 7a, the meter does not meet specification.

Table 7a Frequency Test:

Step

Range

Source

Level

 

 

 

 

1

100Hz

9.0Hz

3V rms

 

 

 

 

2

100Hz

90Hz

3V rms

 

 

 

 

3

1kHz

900Hz

3V rms

 

 

 

 

4

10kHz

9kHz

3V rms

 

 

 

 

5

100kHz

90kHz

3V rms

 

 

 

 

6

1MHz

900kHZ

3V rms

 

 

 

 

Reading

8.987 to 9.013

89.87to 90.13

898.7to 901.3

8.987 to 9.013

89.87to 90.13 898.7to 901.3

Testing the Duty Cycle Function (for 61-340 / 61-342)

To verify the accuracy of the meter’s frequency function, do the following:

1. Connect the calibrator to the

and COM inputs on the meter.

2.Note: The accuracy of the calibrator’s frequency function must be appropriate for the specified accuracy of the meter.

3.Set the rotary switch to .

a. Press the Hz/DUTY button to select the DUTY CYCLE function

4.Set the calibrator or function generator for the square wave voltage and frequency as Table 7b.

5.Compare the reading on the meter display with the display reading shown in Table 7b.

6.If the display reading falls outside of the range shown in Table 7b, the meter does not meet specification.

Table 7b Duty Cycle Test:

Step

Range

Source

Level

 

 

 

 

1

1kHz

1kHz

3V rms

 

 

 

 

Reading

48.3% to 51.7%

Form Number TM61340-2

Rev 2 November 2007

Image 22
Contents Ideal INDUSTRIES, INC. Technical Manual Table of Contents Table A. The Symbols IntroductionSafety Specifications General specifications Characteristics DescriptionInput Impedance 10MΩ Range Resolution Accuracy Over voltage ProtectionRange Resolution Accuracy Input Protection Frequency response 40~400HzRange Resolution Accuracy Range Resolution Accuracy Max. Test Max. Open Current Overload Protection Range Resolution Sensitivity AccuracyRange 15 Hz/DUTY 8 14 REL16 MAX/MIN Environmental characteristics Description Physical and Environmental Characteristics DescriptionCertifications and compliances Frequency Range 40 ~ 1KHz Required EquipmentAC Voltage Range 0 ~ 750V AC Accuracy ±2% DC Voltage Range 0 ~ 1000V DCReplacing the Battery Opening the Meter CaseFuse Replacement Replacing FusesTesting the Display LCD Graphics 61-340Testing the Voltage Function for 61-340 Step Range Input ReadingStep Range Input Frequency Reading Testing the DC microamperes Function for 61-340Testing the DC milli amperes Function for 61-340 Step Range SourceTesting the AC microampere Function for 61-340 Testing the DC ampere Function for 61-340Step Range Source Frequency Reading Testing the AC a function for 61-340 Testing the AC milliampere function for 61-340andStep Source Frequency Reading Testing the Capacitance Function for 61-340 Testing the Resistance Function for 61-340Step Range Source Reading Checking the Diode Test Function for 61-340 Checking the Continuity Function for 61-340Testing the Frequency Function for 61-340 Testing the Duty Cycle Function for 61-340Testing the Temperature Function for 61-340 DC V Calibration Adjust VR1 AC V Calibration Adjust VR2 ModelTemp Calibration Adjust VR3