Page 17
Testing the AC milliampere function: (for 61-340and 61-342)
1.Connect the calibrator to the and COM inputs on the meter.
2.Turn the rotary switch to .
a.Press the SELECT button to select the AC function
b.Press the REL button to make the display is zero
3.Apply the inputs for steps
4.For each input, compare the reading on the meter display to the reading for your meter in Table 4b
5.If the display reading falls outside of the range shown in Table 4b, the meter does not meet specification.
Table 4b ACmA Test:
Step | Range | Source | Frequency | Reading |
1 | 40 mA | 30.00mA | 60Hz | 29.35 to 30.65 |
|
|
|
|
|
2 | 40 mA | 30.00mA | 400Hz | 29.35 to 30.65 |
|
|
|
|
|
3 | 400 mA | 300.0mA | 60Hz | 293.5 to 306.5 |
|
|
|
|
|
4 | 400 mA | 300.0mA | 400Hz | 293.5 to 306.5 |
|
|
|
|
|
Testing the AC A function (for 61-340 / 61-342)
1.Connect the calibrator to the 10A and COM inputs on the meter.
2.Turn the rotary switch to .
a.Press the SELECT button to select the AC function
b.Press the REL button to make the display is zero
3.Apply the inputs for steps
4.For each input, compare the reading on the meter display to the reading for your meter in Table 4c
5.If the display reading falls outside of the range shown in Table 4c, the meter does not meet specification.
Table 4c AC A Test:
Step |
| Source | Frequency | Reading |
1 | 4A | 3.000A | 60Hz | 2.920 to 3.080 |
|
|
|
|
|
2 | 4A | 3.000A | 400Hz | 2.920 to 3.080 |
|
|
|
|
|
3 | 10A | 10A | 60Hz | 9.70 to 10.30 |
|
|
|
|
|
4 | 10A | 10A | 400Hz | 9.70 to 10.30 |
|
|
|
|
|
Form Number | Rev 2 November 2007 |