Page 22
Calibration Procedure
Recalibrate your meter:
It is recommended that the multimeter be calibrated once each year.
1.Perform calibration at an ambient temperature of 23°C±2°C and a relative humidity of <70% Disconnect the test leads and turn the meter off. Remove the test leads from the front terminals.
2.Position the meter face down. Remove the battery cover screw and the 2 bottom case cover screws.
3.Lift the end of the bottom case cover until it gently unsnaps from the case top at the end nearest the LCD.
(A)DC V Calibration (Adjust VR1)
1. Set the rotary switch to position “ ” for the DC Volts function.
2. Set the output of DC calibrator for 300.0mV and connect to | and COM input terminals on |
meter.
3.Using a small
4.Disconnect the DC calibrator from the meter.
(B)AC V Calibration (Adjust VR2) Model 61-340:
1. | Set the rotary switch to position. “ | ” for the AC Volts function. |
|
2. | Set the output of the calibrator for 3.000V at 60Hz and connect to | and COM input | |
| terminals on meter. |
|
|
3.Using a small
4.Disconnect the AC calibrator from the meter.
Model |
|
| |
1. | Set the rotary switch to position. “ | ” and press SELECT button to | select AC A function, |
| then press RANGE button to select | 10A range . |
|
2. | Using a small | ||
3. | Press RANGE button to select 4A | range and adjust VR5 until the display reads same as the | |
| digits at AC 10A range ( 0 to 3 digits) |
| |
4. | Set the rotary switch to position. “ | ” for the AC Volts Function. |
|
5. | Set the output of the calibrator for 3.000V at 60Hz. and connect to | and COM input | |
| terminals on meter. |
|
|