Replacing Fuses

Warning

To avoid electrical shock, remove the test leads and any input signals before replacing the battery or fuses. To prevent damage or injury, INSTALL ONLY quick acting fuses with the following Volt/Amp current interrupt rating:

F340: 500mA, 250V Fast Blow Fuse

F342: 10A, 250V Fast Blow Fuse

Fuse Replacement

The 61-340 and 61-342 are fused in both the μA /mA input and 10 amperes input ports.

μA /mA input port is fused by a F340: 500mA, 250V FAST BLOW fuse, 10 A input port is fused by

F342: 10A, 250V FAST BLOW fuse

1.Remove the rubber holster

2.Remove the screw on the battery cover and open the battery cover.

3.Remove the two screws at the bottom of back cover and a screw under the support stand.

4.Take out the old battery, and remove the back case cover.

5.* Replace the defective fuse with the standard rated fuse for this meter

6.Return the back case cover, reinsert the case screws.

a. It is recommended you replace the old battery with new battery.

7.Return the battery cover and reinstall the battery cover screw.

8.Replace the rubber boot.

Use only a fuse with the amperage, interrupt voltage, and speed rating specified.

Cleaning

Warning

To avoid electrical shock or damage to the meter, never allow water inside the case. To avoid damaging the meter’s housing, never apply solvents to the meter.

Form Number TM61340-2

Rev 2 November 2007

Image 14
Contents Ideal INDUSTRIES, INC. Technical Manual Table of Contents Table A. The Symbols IntroductionSafety Specifications General specifications Characteristics DescriptionInput Impedance 10MΩ Range Resolution Accuracy Over voltage ProtectionRange Resolution Accuracy Input Protection Frequency response 40~400HzRange Resolution Accuracy Range Resolution Accuracy Max. Test Max. Open Current Range Range Resolution Sensitivity AccuracyOverload Protection 16 MAX/MIN 8 14 REL15 Hz/DUTY Certifications and compliances Physical and Environmental Characteristics DescriptionEnvironmental characteristics Description Frequency Range 40 ~ 1KHz Required EquipmentAC Voltage Range 0 ~ 750V AC Accuracy ±2% DC Voltage Range 0 ~ 1000V DCReplacing the Battery Opening the Meter CaseFuse Replacement Replacing FusesTesting the Display LCD Graphics 61-340Testing the Voltage Function for 61-340 Step Range Input ReadingStep Range Input Frequency Reading Testing the DC microamperes Function for 61-340Testing the DC milli amperes Function for 61-340 Step Range SourceStep Range Source Frequency Reading Testing the DC ampere Function for 61-340Testing the AC microampere Function for 61-340 Step Source Frequency Reading Testing the AC milliampere function for 61-340andTesting the AC a function for 61-340 Step Range Source Reading Testing the Resistance Function for 61-340Testing the Capacitance Function for 61-340 Checking the Diode Test Function for 61-340 Checking the Continuity Function for 61-340Testing the Frequency Function for 61-340 Testing the Duty Cycle Function for 61-340Testing the Temperature Function for 61-340 DC V Calibration Adjust VR1 AC V Calibration Adjust VR2 ModelTemp Calibration Adjust VR3