IDEAL INDUSTRIES Testing the Voltage Function for 61-340, Step Range Input Reading

Page 16

Page 14

Testing the Voltage Function (for 61-340 / 61-342)

To verify accuracy in the AC and DC voltage ranges, do the following:

1. Turn the rotary switch to “ ” position.

2. Connect the calibrator to the

and COM inputs on the meter.

3.Set the calibrator for the voltage from step 1 to 7 in Table 1.

4.Compare the reading on the meter display with the display reading shown in Table 1.

5.If the display reading falls outside of the range shown in Table 1, the meter does not meet specification.

Table 1 DC Voltage Test:

 

Step

Range

 

Input

Reading

 

1

400mV

-300.0mV

-297.3 to -302.7

 

 

 

 

 

 

 

2

400mV

 

4.0mV

3.6 to 4.4

 

 

 

 

 

 

 

3

400mV

 

300.0mV

297.3 to 302.7

 

 

 

 

 

 

 

4

4V

 

3.000V

2.975 to 3.025

 

 

 

 

 

 

 

5

40V

 

30.00V

29.75 to 30.25

 

 

 

 

 

 

 

6

400V

 

300.0V

297.5 to 302.5

 

 

 

 

 

 

 

7

1000V

 

1000V

987 to 1013

 

 

 

 

 

 

1. Turn the rotary switch to position.

 

2.Set the calibration for the voltage and frequency from step 1 to 10 in Table 2.

3.Compare the reading on the meter display with the display reading shown in Table 2.

4.If the display reading falls outside of the range shown in Table 2, the meter does not meet specification.

Form Number TM61340-2

Rev 2 November 2007

Image 16
Contents Ideal INDUSTRIES, INC. Technical Manual Table of Contents Table A. The Symbols IntroductionSafety Specifications General specifications Characteristics DescriptionRange Resolution Accuracy Over voltage Protection Range Resolution Accuracy Input ProtectionInput Impedance 10MΩ Frequency response 40~400HzRange Resolution Accuracy Range Resolution Accuracy Max. Test Max. Open Current Overload Protection Range Resolution Sensitivity AccuracyRange 15 Hz/DUTY 8 14 REL16 MAX/MIN Environmental characteristics Description Physical and Environmental Characteristics DescriptionCertifications and compliances Required Equipment AC Voltage Range 0 ~ 750V ACFrequency Range 40 ~ 1KHz Accuracy ±2% DC Voltage Range 0 ~ 1000V DCReplacing the Battery Opening the Meter CaseFuse Replacement Replacing FusesTesting the Display LCD Graphics 61-340Testing the Voltage Function for 61-340 Step Range Input ReadingTesting the DC microamperes Function for 61-340 Testing the DC milli amperes Function for 61-340Step Range Input Frequency Reading Step Range SourceTesting the AC microampere Function for 61-340 Testing the DC ampere Function for 61-340Step Range Source Frequency Reading Testing the AC a function for 61-340 Testing the AC milliampere function for 61-340andStep Source Frequency Reading Testing the Capacitance Function for 61-340 Testing the Resistance Function for 61-340Step Range Source Reading Checking the Diode Test Function for 61-340 Checking the Continuity Function for 61-340Testing the Frequency Function for 61-340 Testing the Duty Cycle Function for 61-340Testing the Temperature Function for 61-340 DC V Calibration Adjust VR1 AC V Calibration Adjust VR2 ModelTemp Calibration Adjust VR3