Page 5
(3b) DC Current
Range | Resolution | Accuracy | Input Protection | |
4.000A | 0.001A | ±(1.5% reading + 5 digits) | 10A, 250V Fast Blow | |
10.00A* | 0.01A | Fuse | ||
|
Overload Protection: A Input: 10A, 250V Fast Blow fuse.
*Caution: Do not make high current measurements on the 10A scale for longer that 15 seconds. This should be followed by a 15 minute cool down period. Exceeding 15 seconds may cause damage to the meter and/or the test leads.
(4a) AC
Range | Resolution | Accuracy | |
400.0µA | 0.1μA | ±(1.5% reading + 5 digits) *1 | |
|
| 40Hz ~ 400Hz | |
4000µA | 1μA | ||
| |||
40.00mA | 10μA | ±(2.0% reading + 5 digits) *1 | |
|
| 40Hz ~ 400Hz | |
400.0mA | 0.1mA | ||
|
Input Protection
500mA, 250V Fast
Blow Fuse
Overload Protection: μA / mA Input: 500mA,250V Fast Blow fuse.
(4b) AC Current
Range | Resolution | Accuracy | Input Protection |
4.000A | 0.001mA | ±(2.5% reading + 5 digits) *1 | 10A, 250V Fast Blow |
10.00A* | 0.01mA | 40Hz ~ 400Hz | Fuse. |
Overload Protection: A Input: 10A (250V) fast blow fuse.
AC Conversion Type: AC conversions are True RMS responding, calibrated to the sine wave input. *1 The specified accuracy is for sine wave at full scale and
up to 2.
*Caution: Do not make high current measurements on the 10A scale for longer that 15 seconds. This should be followed by a 15 minute cool down period. Exceeding 15 seconds may cause damage to the meter and/or the test leads.
Form Number | Rev 2 November 2007 |