Page 5
(3b) DC Current 
Range | Resolution  | Accuracy  | Input Protection  | |
4.000A  | 0.001A  | ±(1.5% reading + 5 digits)  | 10A, 250V Fast Blow  | |
10.00A*  | 0.01A  | Fuse  | ||
  | 
Overload Protection: A Input: 10A, 250V Fast Blow fuse. 
*Caution: Do not make high current measurements on the 10A scale for longer that 15 seconds. This should be followed by a 15 minute cool down period. Exceeding 15 seconds may cause damage to the meter and/or the test leads.
(4a) AC 
Range | Resolution | Accuracy | |
400.0µA  | 0.1μA  | ±(1.5% reading + 5 digits) *1  | |
  | 
  | 40Hz ~ 400Hz  | |
4000µA  | 1μA  | ||
  | |||
40.00mA  | 10μA  | ±(2.0% reading + 5 digits) *1  | |
  | 
  | 40Hz ~ 400Hz  | |
400.0mA  | 0.1mA  | ||
  | 
Input Protection
500mA, 250V Fast
Blow Fuse
Overload Protection: μA / mA Input: 500mA,250V Fast Blow fuse. 
(4b) AC Current 
Range | Resolution  | Accuracy  | Input Protection  | 
4.000A  | 0.001mA  | ±(2.5% reading + 5 digits) *1  | 10A, 250V Fast Blow  | 
10.00A*  | 0.01mA  | 40Hz ~ 400Hz  | Fuse.  | 
Overload Protection: A Input: 10A (250V) fast blow fuse.
AC Conversion Type: AC conversions are True RMS responding, calibrated to the sine wave input. *1 The specified accuracy is for sine wave at full scale and 
up to 2.
*Caution: Do not make high current measurements on the 10A scale for longer that 15 seconds. This should be followed by a 15 minute cool down period. Exceeding 15 seconds may cause damage to the meter and/or the test leads.
Form Number   | Rev 2 November 2007  |