IDEAL INDUSTRIES 61-340 technical manual Opening the Meter Case, Replacing the Battery

Page 13

Page 11

Basic Maintenance

Warning

To avoid shock, remove the test leads and any input signals before opening the case or replacing the battery or fuses.

Opening the Meter Case

Caution

To avoid unintentional shock circuit, always place the uncovered meter assembly on a protective surface. When the case of the meter is open, circuit connections are exposed.

1.Disconnect test leads from any live source, turn the rotary switch to OFF, and remove the test leads from front terminals.

2.For battery replacement, follow instructions under Replacing the Battery section

3.For fuse replacement, follow instructions under Fuse Replacement section.

Replacing the Battery

The 61-340/61-342 meter are both powered by 9V, types are, NEDA 1604, JIS006P, IEC 6F22

1.Remove the rubber holster

2.Remove the screw on the battery cover and open the battery cover

3.Remove the old battery and install the new battery

4.Return the battery cover and reinstall the battery cover screw.

5.Replace the rubber boot.

Battery and Fuse Replacement

Figure 1

Form Number TM61340-2

Rev 2 November 2007

Image 13
Contents Ideal INDUSTRIES, INC. Technical Manual Table of Contents Introduction Table A. The SymbolsSafety General specifications Characteristics Description SpecificationsRange Resolution Accuracy Input Protection Range Resolution Accuracy Over voltage ProtectionInput Impedance 10MΩ Frequency response 40~400HzRange Resolution Accuracy Range Resolution Accuracy Max. Test Max. Open Current Overload Protection Range Resolution Sensitivity AccuracyRange 15 Hz/DUTY 8 14 REL16 MAX/MIN Environmental characteristics Description Physical and Environmental Characteristics DescriptionCertifications and compliances AC Voltage Range 0 ~ 750V AC Required EquipmentFrequency Range 40 ~ 1KHz Accuracy ±2% DC Voltage Range 0 ~ 1000V DCOpening the Meter Case Replacing the BatteryReplacing Fuses Fuse ReplacementLCD Graphics 61-340 Testing the DisplayStep Range Input Reading Testing the Voltage Function for 61-340Testing the DC milli amperes Function for 61-340 Testing the DC microamperes Function for 61-340Step Range Input Frequency Reading Step Range SourceTesting the AC microampere Function for 61-340 Testing the DC ampere Function for 61-340Step Range Source Frequency Reading Testing the AC a function for 61-340 Testing the AC milliampere function for 61-340andStep Source Frequency Reading Testing the Capacitance Function for 61-340 Testing the Resistance Function for 61-340Step Range Source Reading Checking the Continuity Function for 61-340 Checking the Diode Test Function for 61-340Testing the Duty Cycle Function for 61-340 Testing the Frequency Function for 61-340Testing the Temperature Function for 61-340 AC V Calibration Adjust VR2 Model DC V Calibration Adjust VR1Temp Calibration Adjust VR3