Page 4
Measurement Characteristics
Accuracy is ±(% reading + number of digits) at 23°C ± 5°C, less than 75% R.H.
(1) DC Volts(for 
  | Range | Resolution | Accuracy | Over voltage | 
  | protection | |||
  | 
  | 
  | 
  | |
  | 
  | 
  | 
  | 
  | 
  | 400.0mV  | 0.1mV  | ±(0.8% reading + 3 digits)  | 
  | 
  | 4.000V  | 1mV  | 
  | 
  | 
  | 40.00V  | 10mV  | ±(0.8% reading + 1 digits)  | 1000VDC or 750VAC  | 
  | 
  | 
  | 
  | 
  | 
  | 400.0V  | 100mV  | 
  | 
  | 
  | 1000V  | 1V  | ±(1.0% reading + 3 digits)  | 
  | 
Input Impedance: 10MΩ | 
  | 
  | ||
(2) AC Volts   | 
  | 
  | ||
  | 
  | 
  | 
  | 
  | 
  | Range  | Resolution  | Accuracy  | Over voltage  | 
  | protection  | |||
  | 
  | 
  | 
  | |
  | 400.0mV  | 0.1mV  | ±(1.2%+8)  | 
  | 
  | 4.000V  | 1mV  | 
  | 
  | 
  | 
  | 
  | 
  | 1000VDC or 750VAC  | 
  | 40.00V  | 10mV  | 1.0% + 5  | |
  | 
  | 
  | 
  | 
  | 
  | 400.0V  | 100mV  | 
  | 
  | 
  | 750V  | 1V  | ±(1.2%+5)  | 
  | 
Input Impedance: 10MΩ
AC Conversion Type: 
Crest Factor: C.F. = Peak/RMS
For 
Frequency response: 40~400Hz
(3a) DC   | 
  | |||
  | 
  | 
  | 
  | 
  | 
  | Range | Resolution | Accuracy | Input Protection | 
  | 400.0µA  | 0.1μA  | ±(0.8% reading + 2 digits)*  | 
  | 
  | 4000µA  | 1μA  | 500mA, 250V Fast  | |
  | 
  | |||
  | 40.00mA  | 10μA  | ±(0.8% reading + 3 digits)*  | Blow Fuse  | 
  | 400.0mA  | 0.1mA  | 
  | |
  | 
  | 
  | ||
Overload Protection: mA Input: 500mA, 250V Fast Blow fuse. 
Form Number   | Rev 2 November 2007  |