IDEAL INDUSTRIES 61-340 Range Resolution Accuracy Over voltage Protection, Input Impedance 10MΩ

Page 6

Page 4

Measurement Characteristics

Accuracy is ±(% reading + number of digits) at 23°C ± 5°C, less than 75% R.H.

(1) DC Volts(for 61-340 / 61-342)

 

Range

Resolution

Accuracy

Over voltage

 

protection

 

 

 

 

 

 

 

 

 

 

400.0mV

0.1mV

±(0.8% reading + 3 digits)

 

 

4.000V

1mV

 

 

 

40.00V

10mV

±(0.8% reading + 1 digits)

1000VDC or 750VAC

 

 

 

 

 

 

400.0V

100mV

 

 

 

1000V

1V

±(1.0% reading + 3 digits)

 

Input Impedance: 10MΩ

 

 

(2) AC Volts (61-340/61-342)

 

 

 

 

 

 

 

 

Range

Resolution

Accuracy

Over voltage

 

protection

 

 

 

 

 

400.0mV

0.1mV

±(1.2%+8)

 

 

4.000V

1mV

 

 

 

 

 

 

1000VDC or 750VAC

 

40.00V

10mV

1.0% + 5

 

 

 

 

 

 

400.0V

100mV

 

 

 

750V

1V

±(1.2%+5)

 

Input Impedance: 10MΩ

AC Conversion Type: 61-340:Average sensing rms indication calibrated to the sine wave input. 61-342:AC conversion is True RMS responding, calibrated to a sinusoidal waveform

Crest Factor: C.F. = Peak/RMS

For non-sinusoidal waveform, C.F. > 2 add ±1% to accuracy,

Frequency response: 40400Hz

(3a) DC micro-amp and milli-amps (for 61-340 / 61-342)

 

 

 

 

 

 

 

Range

Resolution

Accuracy

Input Protection

 

400.0µA

0.1μA

±(0.8% reading + 2 digits)*

 

 

4000µA

1μA

500mA, 250V Fast

 

 

 

40.00mA

10μA

±(0.8% reading + 3 digits)*

Blow Fuse

 

400.0mA

0.1mA

 

 

 

 

Overload Protection: mA Input: 500mA, 250V Fast Blow fuse. (61-340 / 61-342)

Form Number TM61340-2

Rev 2 November 2007

Image 6
Contents Ideal INDUSTRIES, INC. Technical Manual Table of Contents Table A. The Symbols IntroductionSafety Specifications General specifications Characteristics DescriptionInput Impedance 10MΩ Range Resolution Accuracy Over voltage ProtectionRange Resolution Accuracy Input Protection Frequency response 40~400HzRange Resolution Accuracy Range Resolution Accuracy Max. Test Max. Open Current Range Resolution Sensitivity Accuracy Overload ProtectionRange 8 14 REL 15 Hz/DUTY16 MAX/MIN Physical and Environmental Characteristics Description Environmental characteristics DescriptionCertifications and compliances Frequency Range 40 ~ 1KHz Required EquipmentAC Voltage Range 0 ~ 750V AC Accuracy ±2% DC Voltage Range 0 ~ 1000V DCReplacing the Battery Opening the Meter CaseFuse Replacement Replacing FusesTesting the Display LCD Graphics 61-340Testing the Voltage Function for 61-340 Step Range Input ReadingStep Range Input Frequency Reading Testing the DC microamperes Function for 61-340Testing the DC milli amperes Function for 61-340 Step Range SourceTesting the DC ampere Function for 61-340 Testing the AC microampere Function for 61-340Step Range Source Frequency Reading Testing the AC milliampere function for 61-340and Testing the AC a function for 61-340Step Source Frequency Reading Testing the Resistance Function for 61-340 Testing the Capacitance Function for 61-340Step Range Source Reading Checking the Diode Test Function for 61-340 Checking the Continuity Function for 61-340Testing the Frequency Function for 61-340 Testing the Duty Cycle Function for 61-340Testing the Temperature Function for 61-340 DC V Calibration Adjust VR1 AC V Calibration Adjust VR2 ModelTemp Calibration Adjust VR3