IDEAL INDUSTRIES technical manual Checking the Diode Test Function for 61-340

Page 21

Page 19

Table 6 Capacitance Test:

Step

Range

Source

Reading

1

40nF

10.000nF

9.60 to 10.40

 

 

 

 

2

40nF

30.00nF

29.00 to 31.00

 

 

 

 

3

400nF

300.0nF

290.5 to 309.5

 

 

 

 

4

4µF

3.000µF

2.905 to 3.095

 

 

 

 

5

40.00µF

30.00µF

29.05 to 30.95

 

 

 

 

6

100.0µF

100.0µF

79.5 to 120.5

 

 

 

 

7

1000µF

1000µF

795 to 1205

 

 

 

 

Checking the Diode Test Function (for 61-340 / 61-342)

To check the diode test function, do the following:

1.Connect the DVM (Digital Voltage Meter) to the and COM inputs on the meter.

2.Turn the meter’s rotary switch to .

 

Press the SELECT button to select the

function

 

The DVM display should display OL

 

3.

Insert a Si diode with correct polarity “+” and “-“

 

 

The meter display should read within 0.5~0.7Vdc

4.

Reverse polarity on the Si diode

 

 

The meter display should read OL

 

.

Checking the Continuity Function (for 61-340 / 61-342)

To check the diode test function, do the following:

1.Connect the DVM (Digital Voltage Meter) to the and COM inputs on the meter.

2.Turn the meter’s rotary switch to .

a)Press the SELECT button to select the function

b)The DVM display should display OL

3.Insert a 50Ω resistor

a)The meter inner beeper will sound

Form Number TM61340-2

Rev 2 November 2007

Image 21
Contents Ideal INDUSTRIES, INC. Technical Manual Table of Contents Introduction Table A. The SymbolsSafety General specifications Characteristics Description SpecificationsRange Resolution Accuracy Input Protection Range Resolution Accuracy Over voltage ProtectionInput Impedance 10MΩ Frequency response 40~400HzRange Resolution Accuracy Range Resolution Accuracy Max. Test Max. Open Current Range Resolution Sensitivity Accuracy Overload ProtectionRange 8 14 REL 15 Hz/DUTY16 MAX/MIN Physical and Environmental Characteristics Description Environmental characteristics DescriptionCertifications and compliances AC Voltage Range 0 ~ 750V AC Required EquipmentFrequency Range 40 ~ 1KHz Accuracy ±2% DC Voltage Range 0 ~ 1000V DCOpening the Meter Case Replacing the BatteryReplacing Fuses Fuse ReplacementLCD Graphics 61-340 Testing the DisplayStep Range Input Reading Testing the Voltage Function for 61-340Testing the DC milli amperes Function for 61-340 Testing the DC microamperes Function for 61-340Step Range Input Frequency Reading Step Range SourceTesting the DC ampere Function for 61-340 Testing the AC microampere Function for 61-340Step Range Source Frequency Reading Testing the AC milliampere function for 61-340and Testing the AC a function for 61-340Step Source Frequency Reading Testing the Resistance Function for 61-340 Testing the Capacitance Function for 61-340Step Range Source Reading Checking the Continuity Function for 61-340 Checking the Diode Test Function for 61-340Testing the Duty Cycle Function for 61-340 Testing the Frequency Function for 61-340Testing the Temperature Function for 61-340 AC V Calibration Adjust VR2 Model DC V Calibration Adjust VR1Temp Calibration Adjust VR3