Page 19
Table 6 Capacitance Test:
Step | Range | Source  | Reading  | 
1  | 40nF  | 10.000nF  | 9.60 to 10.40  | 
  | 
  | 
  | 
  | 
2  | 40nF  | 30.00nF  | 29.00 to 31.00  | 
  | 
  | 
  | 
  | 
3  | 400nF  | 300.0nF  | 290.5 to 309.5  | 
  | 
  | 
  | 
  | 
4  | 4µF  | 3.000µF  | 2.905 to 3.095  | 
  | 
  | 
  | 
  | 
5  | 40.00µF  | 30.00µF  | 29.05 to 30.95  | 
  | 
  | 
  | 
  | 
6  | 100.0µF  | 100.0µF  | 79.5 to 120.5  | 
  | 
  | 
  | 
  | 
7  | 1000µF  | 1000µF  | 795 to 1205  | 
  | 
  | 
  | 
  | 
Checking the Diode Test Function (for 61-340 /  61-342) 
To check the diode test function, do the following:
1.Connect the DVM (Digital Voltage Meter) to the 
 and COM inputs on the meter.
2.Turn the meter’s rotary switch to 
.
  | Press the SELECT button to select the  | function  | 
  | The DVM display should display OL  | 
  | 
3.  | Insert a Si diode with correct polarity “+” and   | 
  | 
  | The meter display should read within 0.5~0.7Vdc  | |
4.  | Reverse polarity on the Si diode  | 
  | 
  | The meter display should read OL  | 
  | 
.
Checking the Continuity Function (for 61-340 /  61-342) 
To check the diode test function, do the following:
1.Connect the DVM (Digital Voltage Meter) to the 
 and COM inputs on the meter.
2.Turn the meter’s rotary switch to 
.
a)Press the SELECT button to select the 
 function
b)The DVM display should display OL
3.Insert a 50Ω resistor
a)The meter inner beeper will sound
Form Number   | Rev 2 November 2007  |