HP 8753ET manual HP 8753ES Option 011 specifications, Test port output characteristics

Page 12

HP 8753ES Option 011 specifications

Test port output characteristics

Frequency characteristics

Range

300 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Resolution

1 Hz

Stability

typically ±7.5 ppm 0˚ to 55 ˚C

 

typically ±3 ppm/year

Option 1D5

typically ±0.05 ppm 0˚ to 55 ˚C

 

typically ±0.5 ppm/year

Accuracy

±10 ppm at 25 ˚C ± 5 ˚C

Power range

-5 to +20 dBm

Option 006

-5 to +18 dBm

Resolution

0.01 dB

Level accuracy1,2

±1.0 dB

Level linearity1,2,5

±0.25 dB, -5 to +15 dBm

 

±0.5 dB, +15 to +20 dBm

Impedance

50 Ω nominal

300 kHz to 3 GHz2 >16 dB RL (<1.38 SWR)

300 kHz to 6 GHz

>14 dB RL (<1.50 SWR)

Spectral purity

 

2nd harmonic3

<-40 dBc at +10 dBm6

 

<-50 dBc at 0 dBm6

Option 002

<-25 dBc at max power

3rd harmonic4

<-40 dBc at +10 dBm6

 

<-50 dBc at 0 dBm6

Option 002

<-25 dBc at max power

Nonharmonic spurious6

Mixer related

<-30 dBc at max power

 

<-55 dBc at -10 dBm

Test port input characteristics Option 011

Frequency range

300 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Average noise level2

 

 

50 kHz to 3 GHz

<-90 dBm (3 kHz BW)

50 kHz to 3 GHz

<-110 dBm (10 Hz BW)

50 kHz to 3 GHz

<-120 dBm (10 Hz BW) typical

3 to 6 GHz

<-85 dBm (3 kHz BW)

3 to 6 GHz

<-105 dBm (10 Hz BW)

3 to 6 GHz

<-115 dBm (10 Hz BW) typical

Maximum input level

 

0 dBm

Damage level

20 dBm or 25 VDC

Impedance:

50 ohms nominal

300 kHz to 2 MHz

20 dB RL

2 MHz to 1.3 GHz

24 dB RL

1.3 GHz to 3 GHz

19 dB RL

Option 006

 

15 dB RL6

3 GHz to 6 GHz

Frequency response

 

 

300 kHz to 3 GHz

±1.0 dB,

3 GHz to 6 GHz

 

±2.0 dB

Harmonics (Option 002)

 

2nd harmonic3

<-15 dBc at 0 dBm

 

<-30 dBc at -10 dBm6

 

<-45 dBc at -30 dBm6

3rd harmonic4

<-30 dBc at 0 dBm

 

<-50 dBc at -10 dBm6

 

<-50 dBc at -30 dBm6

Harmonic measurement accuracy

16 MHz to 3 GHz

 

±1.5 dB

3 GHz to 6 GHz Option 006 ±3 dB

Harmonic measurement dynamic range

-40 dBc6

(output = -10 dBm,

 

input <-15 dBm)

1. +10 dBm output power for the HP 8753ES Option 011.

2. Typical below 300 kHz.

3. 16 MHz to 3 GHz.

4. 16 MHz to 2 GHz.

5. For HP 8753ES Option 011 and Option 006, linearity is specified for the ranges of -5 to +13 dBm and +13 to +18 dBm.

6. Typical performance.

12

Image 12
Contents Specifications Table of contents Measurement port characteristics Definitions and test conditionsDynamic range Measurement uncertaintySystem performance summaries HP 8753ES 50-ohm systemsMm test ports Network analyzer HP 8753ESType-N test ports Calibration kit HP 85032B Test-port cables HP 11857DCalibration kit HP 85033D Test-port cables HP 11857D Frequency rangeHP 8753ES 75-ohm systems Network analyzerCalibration kit HP 85036B Test-port cables HP 11857B Type-F test ports Calibration kit HP 85039B Test-port cables HP 11857B Measurement port characteristics KHz to 16 MHz 100 dB MHz to 3 GHz 110 dB GHz to 6 GHz 105 dBHP 8753ES specifications Test-port output characteristics7Test port input characteristics Characteristics Phase characteristicsCW time sweep only KHz to 6 GHzHP 8753ES Option 011 specifications Test port output characteristicsTest port input characteristics Option KHz to 3 GHz To -35 dBm GHz to 6 GHz To -30 dBm 20 kHzFrequency CW time sweep onlyHP 8753ET specifications Test-port output characteristicsKHz to 3 GHz Option 006 300 kHz to 6 GHzTest-port input characteristics Magnitude characteristics±500 dB ±180˚HP 8753ET/ES supplemental characteristics Group delay characteristicsMeasurement Source controlData accuracy enhancement Data averagingDisplay control Trace functions StorageData hardcopy System capabilitiesRemote programming Test sequencesTime-domain Option This stimulus is also used to measure low-pass devicesGeneral characteristics Options Test set specifications for Option HP 85046A/B SpecificationsHP 85046A/B S-parameter test sets Specifications HP 85047A S-parameter test set+20 dBm +30 dBmAccessories HP 85039B type-F calibration kit Specifications HP 85038 family of calibration kitsHP 11906B 7-16 to 50-Ohm type-N adapter kit HP 11906C 7-16 to 7-mm adapter kitVerification kits Test-port return cablesOhm accessory kits RF limiterHP 85024A high-frequency probe Dimensions WeightHP 8347A RF amplifier