HP 8753ES Option 011 specifications
Test port output characteristics
Frequency characteristics
Range | 300 kHz to 3 GHz |
Option 006 | 30 kHz to 6 GHz |
Resolution | 1 Hz |
Stability | typically ±7.5 ppm 0˚ to 55 ˚C |
| typically ±3 ppm/year |
Option 1D5 | typically ±0.05 ppm 0˚ to 55 ˚C |
| typically ±0.5 ppm/year |
Accuracy | ±10 ppm at 25 ˚C ± 5 ˚C |
Power range | |
Option 006 | |
Resolution | 0.01 dB |
Level accuracy1,2 | ±1.0 dB |
Level linearity1,2,5 | ±0.25 dB, |
| ±0.5 dB, +15 to +20 dBm |
Impedance | 50 Ω nominal |
300 kHz to 3 GHz2 >16 dB RL (<1.38 SWR) | |
300 kHz to 6 GHz | >14 dB RL (<1.50 SWR) |
Spectral purity |
|
2nd harmonic3 | |
| |
Option 002 | |
3rd harmonic4 | |
| |
Option 002 | |
Nonharmonic spurious6 | |
Mixer related | |
|
Test port input characteristics Option 011
Frequency range | 300 kHz to 3 GHz | |
Option 006 | 30 kHz to 6 GHz | |
Average noise level2 |
|
|
50 kHz to 3 GHz | ||
50 kHz to 3 GHz | ||
50 kHz to 3 GHz | ||
3 to 6 GHz | ||
3 to 6 GHz | ||
3 to 6 GHz | ||
Maximum input level |
| 0 dBm |
Damage level | 20 dBm or 25 VDC | |
Impedance: | 50 ohms nominal | |
300 kHz to 2 MHz | ≥20 dB RL | |
2 MHz to 1.3 GHz | ≥24 dB RL | |
1.3 GHz to 3 GHz | ≥19 dB RL | |
Option 006 |
| ≥15 dB RL6 |
3 GHz to 6 GHz | ||
Frequency response |
|
|
300 kHz to 3 GHz | ±1.0 dB, | |
3 GHz to 6 GHz |
| ±2.0 dB |
Harmonics (Option 002) |
| |
2nd harmonic3 | ||
| ||
| ||
3rd harmonic4 | ||
| ||
| ||
Harmonic measurement accuracy | ||
16 MHz to 3 GHz |
| ±1.5 dB |
3 GHz to 6 GHz Option 006 ±3 dB | ||
Harmonic measurement dynamic range | ||
| (output = | |
| input |
1. +10 dBm output power for the HP 8753ES Option 011.
2. Typical below 300 kHz.
3. 16 MHz to 3 GHz.
4. 16 MHz to 2 GHz.
5. For HP 8753ES Option 011 and Option 006, linearity is specified for the ranges of
6. Typical performance.
12