HP 8753ET manual Verification kits, Test-port return cables

Page 26

Verification kits

Measuring known devices other than the standards used in calibration is an easy way to verify the proper operation of an HP 8753ES measurement system. HP offers verification kits which include devices, with data, for verifying the error-corrected measurements of an HP 8753ES and 50-ohm test sets.

HP 85029B 7-mm verification kit

Contains a set of precision 7-mm devices, with data traceable to NIST* for verifying the calibrated perfor- mance of an HP 8753E measurement system.

The HP 85031B 7-mm calibration kit is required for complete verification.

Test-port return cables

Hewlett-Packard offers high-quality RF cables to con- nect the HP 8753ET/ES and test sets to devices under test. These cables offer excellent RF shielding for high- dynamic-range measurements.

HP 11851B 50-ohm type-N RF cable kit

Recommended for use with the HP 11850C/D three- way power splitters. Kit includes three phase-matched 610-mm (24-in.) cables and one 860-mm (34-in.) cable.

Return loss

> 24 dB to 3 GHz

Phase tracking

±4˚ at 1.3 GHz

HP 8120-5639 50-ohm type-N RF cable

A 50-ohm type-N test port cable (both connectors are type-N male).

Return loss

> 30dB to 6 GHz

HP 11857B 75-ohm type-N test port return cables

A pair of 610-mm (24-in.) test port return cables for use with the HP 8753ES or HP 85046B 75-ohm S-para- meter test set.

Return loss

> 24 dB to 2 GHz

Phase tracking

±2˚ at 1.3 GHz

HP 11857D 7-mm test-port return cables

A pair of 610-mm (24-in.) test port return cables for use with the HP 8753ES with the HP 85046A and 85047A S-parameter test sets. These cables can be used with connector types other than 7-mm with the appropriate precision adapters.

Return loss

> 24 dB to 3 GHz

 

> 20 dB to 6 GHz

Phase tracking

±2˚ at 1.3 GHz

HP 11857F type-F test-port return cables Type-F test port cable set (75-ohm). Includes one cable with type-N(male) to type-F(male) connectors and one cable with type-N(male) to type-F(female) connectors.

HP 11850C/D three-way power splitters

 

 

HP 11850C

 

HP 11850D

Impedance

 

50 ohms

 

75 ohms

Frequency range

DC to 3 GHz

 

DC to 2 GHz

Tracking

 

±25 dB, ±3˚

 

±2 dB, ±2.5˚

Equivalent

 

30 dB at 1.3 GHz

30 dB at 1.3 GHz

source match

(ratio or leveling) 20 dB at 3 GHz

20 dB at 2 GHz

Nominal

 

9.5 dB + 1 dB/GHz

7.8 dB

insertion loss

 

 

 

 

Input port match

 

 

 

 

DC to 1.3 GHz

20 dB

 

 

20 dB

1.3 GHz to Fmax

10 dB

 

 

10 dB

Maximum

 

+20 dB

 

 

+20 dB

operating level

 

 

Damage level

 

+30 dB

 

 

+30 dB

RF connectors

 

 

 

 

RF input

50-ohm type-N (f)

50-ohm type-N(f)

All others

50-ohm type-N (f)

75-ohm type-N(f)

Includes

 

Three HP 11852B 50-to 75-ohm

 

 

minimum-loss pads

Recommended accessories

 

 

 

 

 

HP 11851B RF cable kit

HP 11667A 50-ohm power splitter

 

Frequency range

 

DC to 18 GHz

Typical insertion loss

6 dB

 

 

Equivalent source match

 

 

 

DC to 4 GHz

 

26 dB

 

 

4 GHz to 8 GHz

 

21 dB

 

 

8 GHz to 18 GHz

 

17 dB

 

 

Tracking

 

 

 

 

 

(between output arms)

 

 

 

DC to 4 GHz

 

±15 dB

 

 

4 GHz to 8 GHz

 

±2 dB

 

 

8 GHz to 18 GHz

 

±25 dB

 

 

Maximum operating level

±27 dBm

 

Connectors

 

 

50-ohm type-N (f)

Option 001

 

 

 

 

 

RF input

 

 

type-N (m)

Outputs

 

 

type-N (f)

 

Option 002

 

 

 

 

 

RF input

 

 

type-N (f) on RF

Outputs

 

 

precision 7-mm

Dimensions

 

 

46 mm H x 52 mmW x

 

 

 

19 mm D

 

 

 

 

(1.8 x 2.0 x 0.7 in)

Recommended accessories

HP 11851B RF cable kit

HP 11852B 50-ohm to 75-ohm minimum loss pad

Frequency range

DC to 3.0 GHz

Nominal insertion loss

5.7 dB

Return loss

 

300 kHz to 2 GHz

32 dB

2 GHz to 3 GHz

27 dB

Maximum input power

250 mW (+24 dBm)

26

Image 26
Contents Specifications Table of contents Dynamic range Measurement port characteristicsDefinitions and test conditions Measurement uncertaintyMm test ports System performance summariesHP 8753ES 50-ohm systems Network analyzer HP 8753ESType-N test ports Calibration kit HP 85032B Test-port cables HP 11857DCalibration kit HP 85033D Test-port cables HP 11857D Frequency rangeCalibration kit HP 85036B Test-port cables HP 11857B HP 8753ES 75-ohm systemsNetwork analyzer Type-F test ports Calibration kit HP 85039B Test-port cables HP 11857BMeasurement port characteristics KHz to 16 MHz 100 dB MHz to 3 GHz 110 dB GHz to 6 GHz 105 dBTest port input characteristics HP 8753ES specificationsTest-port output characteristics7 CW time sweep only CharacteristicsPhase characteristics KHz to 6 GHzTest port input characteristics Option HP 8753ES Option 011 specificationsTest port output characteristics Frequency KHz to 3 GHz To -35 dBm GHz to 6 GHz To -30 dBm20 kHz CW time sweep onlyKHz to 3 GHz HP 8753ET specificationsTest-port output characteristics Option 006 300 kHz to 6 GHz±500 dB Test-port input characteristicsMagnitude characteristics ±180˚Measurement HP 8753ET/ES supplemental characteristicsGroup delay characteristics Source controlDisplay control Data accuracy enhancementData averaging Data hardcopy Trace functionsStorage System capabilitiesTime-domain Option Remote programmingTest sequences This stimulus is also used to measure low-pass devicesGeneral characteristics Options HP 85046A/B S-parameter test sets Test set specifications for OptionHP 85046A/B Specifications +20 dBm Specifications HP 85047A S-parameter test set +30 dBmAccessories HP 11906B 7-16 to 50-Ohm type-N adapter kit HP 85039B type-F calibration kit SpecificationsHP 85038 family of calibration kits HP 11906C 7-16 to 7-mm adapter kitVerification kits Test-port return cablesHP 85024A high-frequency probe Ohm accessory kitsRF limiter HP 8347A RF amplifier DimensionsWeight