System performance summaries
HP 8753ET
Configuration |
|
|
|
|
Network analyzer | HP 8753ET Standard, | |||
|
| Option 006 or Option 004 | ||
Calibration kit |
| HP 85032B | ||
Test port cable |
| HP part number | ||
System dynamic range |
|
|
| |
300 kHz to 16 MHz | 100 dB |
|
| |
16 MHz to 3 GHz | 110 dB |
|
| |
3 GHz to 6 GHz |
| 105 dB |
|
|
Measurement port characteristics |
|
| ||
|
|
| ||
| Frequency range | |||
Uncorrected 300 | 3.0 | |||
Directivity | 30 dB | 24 dB | 19 dB | |
Source match4 | 25 dB | 20 dB | 14 dB | |
Load match | 24 dB | 19 dB | 16 dB | |
Reflection tracking | ±1.0 dB | ±1.0 dB | ±2.0 dB | |
Transmission tracking | ±1.5 dB | ±1.5 dB | ±2.5 dB | |
Crosstalk | 100 dB | 100 dB | 90 dB |
|
Transmission uncertainty:
Transmission uncertainty: (response calibration)
Reflection uncertainty
Frequency range (continued)
| Corrected | 300 | 1.3 | ||
| Reflection measurements1 |
|
|
| |
Directivity |
| 50 dB | 47 dB | 40 dB | |
Source match |
| 42 dB | 36 dB | 31 dB | |
Load match |
|
|
|
| |
|
| 24 dB | 19 dB | 16 dB | |
| Enhanced reflection cal | 24 dB | 19 dB | 16 dB | |
Reflection tracking |
| ±(0.009 dB | ±(0.019 dB | ±(0.07 dB | |
|
|
| +0.01 dB/˚C) | +0.02 dB/˚C) | +0.03 dB/˚C) |
Transmission measurements2 |
|
| |||
Source match |
|
|
|
| |
| 42 dB | 36 dB | 31 dB | ||
|
| 25 dB | 20 dB | 14 dB | |
Transmission tracking |
|
|
|
| |
| ±(0.006 dB | ±(0.018 dB | ±(0.054 dB | ||
|
|
| +0.01 dB/˚C) | +0.02 dB/˚C) | +0.03 dB/˚C) |
|
| ±(0.033 dB | ±(0.1 dB | ±(0.27 dB | |
|
|
| +0.01 dB/˚C) | +0.02 dB/˚C) | +0.03 dB/˚C) |
1. |
| |
2. |
| |
3. | 2 dB less with Option 004. |
|
4. | Option 004 may degrade uncorrected source match as much as 2 dB. |
|
9 | ||
|
|
|