System performance summaries
HP 8753ES (50-ohm systems)
7-mm test ports
Configuration |
|
Network analyzer | HP 8753ES |
| Standard and Option 006 |
Calibration kit | HP 85031B |
| HP 11857D |
System dynamic range |
|
30 kHz to 50 kHz | 70 dB1 |
50 kHz to 300 kHz | 90 dB1 |
300 kHz to 16 MHz | 100 dB |
16 MHz to 3 GHz | 110 dB |
Option 014 | 108 dB |
3 GHz to 6 GHz | 105 dB |
Option 014 | 103 dB |
Reflection uncertainty
Magnitude
Transmission uncertainty
Magnitude
1.Typical below 300 kHz.
2.Typical performance.
Measurement port characteristics
Frequency range
Corrected 30 | 300 | 1.3 | 3 GHz- 6 GHz | ||
Directivity | 55 dB | 55 dB | 51 dB | 46 dB | |
Source match | 55 dB | 51 dB | 49 dB | 43 dB | |
Load match | 55 dB | 55 dB | 51 dB | 46 dB | |
Reflection | ±(0.001 dB | ±(0.001 dB | ±(0.005 dB | ±(0.020 dB | |
tracking | +0.02 dB/˚C) | +0.01 dB/˚C) | +0.02 dB/˚C) | +0.03 dB/˚C) | |
Transmission | ±(0.008 dB | ±(0.006 dB | ±(0.009 dB | ±(0.021 dB | |
tracking | +0.02 dB/˚C) | +0.01 dB/˚C) | +0.02 dB/˚C) | +0.03 dB/˚C) | |
Uncorrected |
|
|
|
| |
Directivity | 20 dB* | 35 dB | 30 dB | 25 dB | |
Source match | 18 dB** | 16 dB | 16 dB | 14 dB | |
Load match | 18 dB** | 18 dB | 16 dB | 14 dB | |
Reflection | ±2.5 dB | ±1 dB | ±1 dB | ±1.5 dB | |
tracking |
|
|
|
| |
Transmission | ±2.5 dB | ±1 dB | ±1 dB | ±1.5 dB | |
tracking |
|
|
|
| |
Crosstalk | 90 dB*** | 100 dB | 100 dB | 90 dB | |
*15 dB, 30 kHz to 50 kHz | **10 dB, 30 kHz to 50 kHz |
| |||
***60dB, 30 kHz to 50 kHz |
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
Phase
Phase
4