HP 8753ET manual Type-F test ports, Calibration kit HP 85039B Test-port cables HP 11857B

Page 8

System performance summaries

HP 8753ES (75-ohm systems)

Type-F test ports

Configuration

 

Network analyzer

HP 8753E Option 075

Calibration kit

HP 85039B

Test-port cables

HP 11857B

System dynamic range

 

30 kHz to 50 kHz

68 dB1

50 kHz to 300 kHz

90 dB1

300 kHz to 16 MHz

96 dB

16 MHz to 3 GHz

106 dB

Reflection uncertainty

Magnitude

Transmission uncertainty

Magnitude

1.Typical below 300 kHz.

2.Typical performance.

Measurement port characteristics

Data is shown for a Type-F female reflection port and a Type-F male transmission port.

Frequency range

Corrected2

30 kHz-300 kHz

300 kHz-1.3 GHz

1.3 GHz-3 GHz

Directivity

45 dB

45 dB

40 dB

Source match

40 dB

40 dB

30 dB

Load match

45 dB

45 dB

40 dB

Reflection

±(0.060 dB

±(0.060 dB

±(0.024 dB

tracking

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

Transmission

±(0.033 dB

±(0.019 dB

±(0.057 dB

tracking

+0.02 dB/˚C)

+0.01 dB/˚C)

+0.02 dB/˚C)

 

 

 

 

 

 

 

 

 

 

 

 

Phase

Phase

8

Image 8
Contents Specifications Table of contents Measurement port characteristics Definitions and test conditionsDynamic range Measurement uncertaintySystem performance summaries HP 8753ES 50-ohm systemsMm test ports Network analyzer HP 8753ES Type-N test ports Calibration kit HP 85032B Test-port cables HP 11857DCalibration kit HP 85033D Test-port cables HP 11857D Frequency rangeCalibration kit HP 85036B Test-port cables HP 11857B HP 8753ES 75-ohm systemsNetwork analyzer Type-F test ports Calibration kit HP 85039B Test-port cables HP 11857BMeasurement port characteristics KHz to 16 MHz 100 dB MHz to 3 GHz 110 dB GHz to 6 GHz 105 dBTest port input characteristics HP 8753ES specificationsTest-port output characteristics7 Characteristics Phase characteristicsCW time sweep only KHz to 6 GHzTest port input characteristics Option HP 8753ES Option 011 specificationsTest port output characteristics KHz to 3 GHz To -35 dBm GHz to 6 GHz To -30 dBm 20 kHzFrequency CW time sweep onlyHP 8753ET specifications Test-port output characteristicsKHz to 3 GHz Option 006 300 kHz to 6 GHzTest-port input characteristics Magnitude characteristics±500 dB ±180˚HP 8753ET/ES supplemental characteristics Group delay characteristicsMeasurement Source controlDisplay control Data accuracy enhancementData averaging Trace functions StorageData hardcopy System capabilitiesRemote programming Test sequencesTime-domain Option This stimulus is also used to measure low-pass devicesGeneral characteristics Options HP 85046A/B S-parameter test sets Test set specifications for OptionHP 85046A/B Specifications Specifications HP 85047A S-parameter test set+20 dBm +30 dBmAccessories HP 85039B type-F calibration kit Specifications HP 85038 family of calibration kitsHP 11906B 7-16 to 50-Ohm type-N adapter kit HP 11906C 7-16 to 7-mm adapter kitVerification kits Test-port return cablesHP 85024A high-frequency probe Ohm accessory kitsRF limiter HP 8347A RF amplifier DimensionsWeight