HP 8753ET specifications
Test-port input characteristics (continued)
External source mode1 (CW time sweep only)
Frequency range | 300 kHz to 6 GHz |
Power level | 0 to |
Spectral purity |
|
Maximum spurious | |
input | <20 kHz |
Residual FM | |
Typical settling time | 500 ms (automatic) |
| 50 ms (manual) |
Frequency readout | 0.1% (automatic) |
accuracy2 |
|
Input frequency accuracy requirement2 | |
Manual: |
|
Accuracy | (See magnitude and phase |
| characteristics) |
Magnitude characteristics
Dynamic accuracy (10 Hz IF BW)
(300 kHz to 3 GHz)
(3 to 6 GHz)
Display resolution 0.001 dB/division
Marker resolution3 0.001 dB
Trace noise
(0 dBm at transmission port or +5 dBm at reflection port, ratio measurement, 3 kHz BW)
300 kHz to 3 GHz | < 0.006 dB rms | |
3 GHz to 6 GHz | < 0.010 dB rms | |
|
|
|
Reference level | ±500 dB |
Range |
(300 kHz to 3 GHz)
(3 to 6 GHz)
Resolution | 0.001 dB |
Stability2 |
|
300 kHz to 3 GHz | 0.02 dB/ 1 ˚C |
3 GHz to 6 GHz | 0.04 dB/ 1 ˚C |
Phase characteristics
Dynamic accuracy (10 Hz IF BW)
Range | ±180˚ |
Display resolution | 0.01˚/ division |
Marker resolution3 | 0.01˚ |
Trace noise |
|
(0 dBm at transmission port or +5 dBm at reflection port, ratio measurement, 3 kHz BW)
300 kHz to 3 GHz | < 0.038˚ rms |
3 GHz to 6 GHz | < 0.070˚ rms |
Reference level | |
Range | |
Resolution | 0.01˚ |
Stability |
|
30 kHz to 3 GHz | 0.05˚/ ˚C |
3 GHz to 6 GHz | 0.20˚/ ˚C |
Polar characteristics | 10 x |
Range | |
| scale |
Reference | ±500 units |
1.See the HP 8753ET/ES descriptions and options for a functional description. Measurement accuracy is dependent on the stability of the input signal.
2.Typical performance.
3.Marker resolution for magnitude, phase and delay is dependent upon measured value. Resolution is limited to five digits.
15